Usefulness of UF-5000 automatic screening system in UTI diagnosis
Authors:
- 5 July 2023
- journal article
- research article
- Published by Springer Nature in Brazilian Journal of Microbiology
- Vol. 54 (3) , 1803-1808
Abstract
We have evaluated the Sysmex UF-5000 cytometer use in microbiology for the screening of negative urines, looking for cut-off points to detect bacteria and leukocytes. The number of processed urines was 3569, the highest to date in these studies. The best general cut-off point has been 100 bact/μl, giving an area under the ROC curve of 0.868, a sensitivity of 96%, a specificity of 50%, 1.17% of false negatives, and saving 40% of cultures. The PPV and NPV have been 35.5 and 95.4 respectively. The leukocyte count has not been useful. Finally, we have evaluated urine screening usefulness, concluding that in laboratories such as ours (284 urines/working day) or smaller, it is not cost-effective.Keywords
This publication has 22 references indexed in Scilit:
- Evaluation of 3 different rapid automated systems for diagnosis of urinary tract infectionsDiagnostic Microbiology and Infectious Disease, 2012
- Screening for Urinary Tract Infection with the Sysmex UF-1000i Urine Flow CytometerJournal of Clinical Microbiology, 2011
- UF-1000i™ flow cytometry is an effective screening method for urine specimensDiagnostic Microbiology and Infectious Disease, 2011
- Diagnóstico microbiológico de las infecciones del tracto urinarioEnfermedades Infecciosas y Microbiología Clínica, 2011
- Diagnosis of Bacteriuria and Leukocyturia by Automated Flow Cytometry Compared with Urine CultureJournal of Clinical Microbiology, 2010
- Evaluation of the Sysmex UF1000i flow cytometer for ruling out bacterial urinary tract infectionClinica Chimica Acta, 2010
- Evaluation of the Sysmex UF-1000i for the Diagnosis of Urinary Tract InfectionAmerican Journal of Clinical Pathology, 2010
- Urine sediment analysis: Analytical and diagnostic performance of sediMAX® — A new automated microscopy image-based urine sediment analyserClinica Chimica Acta, 2010
- Cutoff values for bacteria and leukocytes for urine flow cytometer Sysmex UF-1000i in urinary tract infectionsDiagnostic Microbiology and Infectious Disease, 2009
- Rapid detection of urinary tract infection – evaluation of flow cytometryClinical Nephrology, 2006