Two-electrode biopotential amplifier with current-driven inputs
- 1 January 2002
- journal article
- research article
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 40 (1) , 122-127
Abstract
A circuit was developed for a differential two-electrode biopotential amplifier. Current sources at the amplifier inputs were controlled by the common-mode voltage. This principle is well known in telephony for interfacing the telephone line with analogue-type phones. A low impedance of about 1 kΩ was obtained between each input and the common point of the circuit. The differential input impedance of 60 MΩ was obtained with the use of precision resistors. Considerable reduction in the common-mode voltages of more than 200 times resulted. The circuit can be useful for biosignal acquisition from subjects in areas of very high electromagnetic fields, where high common-mode voltages could saturate the input amplifier stages.Keywords
This publication has 9 references indexed in Scilit:
- Developments in ECG acquisition, preprocessing, parameter measurement, and recordingIEEE Engineering in Medicine and Biology Magazine, 1998
- Comparative analysis of power-line interference between two- or three-electrode biopotential amplifiersMedical & Biological Engineering & Computing, 1995
- The isolation mode rejection ratio in bioelectric amplifiersIEEE Transactions on Biomedical Engineering, 1991
- High-quality recording of bioelectric eventsMedical & Biological Engineering & Computing, 1990
- Interference-rejection characteristics of biopotential amplifiers: a comparative analysisIEEE Transactions on Biomedical Engineering, 1988
- On the Reduction of Interference Due to Common Mode Voltage in Two-Electrode Biopotential AmplifiersIEEE Transactions on Biomedical Engineering, 1986
- Reduction of Interference Due to Common Mode Voltage in Biopotential AmplifiersIEEE Transactions on Biomedical Engineering, 1983
- Ground-Free ECG Recording with Two ElectrodesIEEE Transactions on Biomedical Engineering, 1980