Correlation of image quality with exposure index and processing protocol in a computed radiography system
- 15 December 2007
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 130 (2) , 162-171
Abstract
The correlation of image quality with the exposure index (EI) and the processing protocol was investigated in a Kodak computed radiography (CR) system using clinical radiographs and a water phantom containing an aluminium and a copper stepwedge. The phantom was exposed to different dose levels and the acquired images were processed using four clinical protocols. The quality of these images was evaluated in terms of image brightness, contrast and noise. In clinical radiographs, there was no straightforward correlation of image quality with EI. In phantom images, higher EI values improved contrast and reduced noise but after a point this improvement does not justify the implied increase in patient dose. Image brightness, contrast and noise were also strongly dependent on the processing protocol. To obtain the images of satisfactory quality with the Kodak CR system, a dose slightly higher than those used in 400 relative speed screen–film systems and a processing protocol designated for the specific radiographic examination are required.Keywords
This publication has 24 references indexed in Scilit:
- Calibrating automatic exposure control devices for digital radiographyPhysics in Medicine & Biology, 2006
- Influence of cassette type on the DQE of CR systemsMedical Physics, 2006
- Application of contrast-to-noise ratio in optimizing beam quality for digital chest radiography: comparison of experimental measurements and theoretical simulationsPhysics in Medicine & Biology, 2006
- Optimization of Dose and Image Quality for Computed Radiography and Digital RadiographyJournal of Digital Imaging, 2006
- Image quality and dose management in digital radiography: a new paradigm for optimisationRadiation Protection Dosimetry, 2005
- Dose optimisation in computed radiographyRadiation Protection Dosimetry, 2005
- Optimising automatic exposure control in computed radiography and the impact on patient doseRadiation Protection Dosimetry, 2005
- Dose–image quality optimisation in digital chest radiographyRadiation Protection Dosimetry, 2005
- The influence of patient thickness and imaging system on patient dose and physical image quality in digital chest imagingRadiation Protection Dosimetry, 2005
- Method of simulating dose reduction for digital radiographic systemsRadiation Protection Dosimetry, 2005