Effects of collimator backscatter in an Elekta linac by Monte Carlo simulation
- 31 August 2009
- journal article
- research article
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 32 (3) , 129-135
Abstract
The effects of radiation backscattered from the secondary collimators into the monitor chamber in an Elekta linac (producing 6 and 10 MV photon beams) are investigated using BEAMnrc Monte Carlo simulations. The degree and effects of this backscattered radiation are assessed by evaluating the changes to the calculated dose in the monitor chamber, and by determining a correction factor for those changes. Additionally, the fluence and energy characteristics of particles entering the monitor chamber from the downstream direction are evaluated by examining BEAMnrc phase-space data. It is shown that the proportion of particles backscattered into the monitor chamber is small (<0.35 %), for all field sizes studied. However, when the backscatter plate is removed from the model linac, these backscattered particles generate a noticeable increase in dose to the monitor chamber (up to ≈2.4 % for the 6 MV beam and up to 4.4 % for the 10 MV beam). With its backscatter plate in place, the Elekta linac (operating at 6 and 10 MV) is subject to negligible variation of monitor chamber dose with field size. At these energies, output variations in photon beams produced by the clinical Elekta linear accelerator can be attributed to head scatter alone. Corrections for field-size-dependence of monitor chamber dose are not necessary when running Monte Carlo simulations of the Elekta linac operating at 6 and 10 MV.Keywords
This publication has 14 references indexed in Scilit:
- Radiotherapy treatment verification using radiological thickness measured with an amorphous silicon electronic portal imaging device: Monte Carlo simulation and experimentPhysics in Medicine & Biology, 2008
- Ecliptic method for the determination of backscatter into the beam monitor chambers in photon beams of medical acceleratorsPhysics in Medicine & Biology, 2007
- Efficient photon beam dose calculations using DOSXYZnrc with BEAMnrcMedical Physics, 2006
- Absolute dose calculations for Monte Carlo simulations of radiotherapy beamsPhysics in Medicine & Biology, 2005
- Output ratio in air for MLC shaped irregular fieldsMedical Physics, 2004
- Monte Carlo modelling of external radiotherapy photon beamsPhysics in Medicine & Biology, 2003
- Monte Carlo calculation of nine megavoltage photon beam spectra using the BEAM codeMedical Physics, 2002
- Modeling photon output caused by backscattered radiation into the monitor chamber from collimator jaws using a Monte Carlo techniqueMedical Physics, 2000
- BEAM: A Monte Carlo code to simulate radiotherapy treatment unitsMedical Physics, 1995
- The effect of collimator backscatter radiation on photon output of linear acceleratorsMedical Physics, 1987