Abstract
The ultrahigh sensitivity of spin-exchange relaxation-free (SERF) magnetometers and comagnetometers plays a crucial role in tests of fundamental physics, biomagnetic measurement, and quantum navigation. The high atomic number density is essential for achieving the SERF regime, making the heating system an indispensable component. However, thermal-magnetic coupling noise generated by the heating system represents a primary source of perturbation for precision measurement based on thermal atomic ensemble. We identify the main sources of thermal-magnetic coupling noise based on a K-Rb-21Ne SERF comagnetometer. By optimizing the operating mode of the heating system, the low-frequency noise at 0.5 Hz is improved by approximately one order of magnitude, thereby enhancing the low-frequency sensitivity of the SERF comagnetometer. Our research advances the long-term stability of atomic sensors and supports their application in probing new physics beyond the Standard Model in the low-energy regime.
Funding Information
  • Quantum Science and Technology-National Science and Technology Major Project (2021ZD0300401)
  • National Natural Science Foundation of China (62473022)
  • National Key Research and Development Program of China (2025YFE0218000)

This publication has 39 references indexed in Scilit: