High‐Definition Zoom Mode: A High Resolution X‐ray Microscope for Neurointerventional Treatment Procedures
Authors:
- Muhammad WaqasMWMuhammad Waqas
- Vernard S. FennellVFVernard S. Fennell
- Gursant S. AtwalGAGursant S. Atwal
- Hussain ShallwaniHSHussain Shallwani
- Daniel R. BednarekDBDaniel R. Bednarek
- Jason M. DaviesJDJason M. Davies
- Kenneth V. SnyderKSKenneth V. Snyder
- Maxim MokinMMMaxim Mokin
- Stephen RudinSRStephen Rudin
- Elad I. LevyELElad I. Levy
Open Access
- 24 July 2019
- journal article
- research article
- Published by Wiley in Journal of Neuroimaging
- Vol. 29 (5) , 565-572
Abstract
BACKGROUND AND PURPOSE: Visualization of structural details of treatment devices during neurointerventional procedures can be challenging. A new true two‐resolution imaging X‐ray detector system features a 194 µm pixel conventional flat‐panel detector (FPD) mode and a 76 µm pixel high‐resolution high‐definition (Hi‐Def) zoom mode in one detector panel. The Hi‐Def zoom mode was developed for use in interventional procedures requiring superior image quality over a small field of view (FOV). We report successful use of this imaging system during intracranial aneurysm treatment in 1 patient with a Pipeline‐embolization device and 1 patient with a low‐profile visualized intramural support (LVIS Blue) device plus adjunctive coiling. METHODS: A guide catheter was advanced from the femoral artery insertion site to the proximity of each lesion using standard FPD mode. Under magnified small FOV Hi‐Def imaging mode, an intermediate catheter and microcatheters were guided to the treatment site, and the PED and LVIS Blue plus coils were deployed. Radiation doses were tracked intraprocedurally. RESULTS: Critical details, including structural changes in the PED and LVIS Blue and position and movement of the microcatheter tip within the coil mass, were more readily apparent in Hi‐Def mode. Skin‐dose mapping indicated that Hi‐Def mode limited radiation exposure to the smaller FOV of the treatment area. CONCLUSIONS: Visualization of device structures was much improved in the high‐resolution Hi‐Def mode, leading to easier, more controlled deployment of stents and coils than conventional FPD mode.Keywords
Funding Information
- National Science Foundation (R01EB2873)
This publication has 8 references indexed in Scilit:
- High-Definition Zoom Mode, a High-Resolution X-Ray Microscope for Neurointerventional Treatment Procedures: A Blinded-Rater Clinical-Utility StudyAmerican Journal of Neuroradiology, 2018
- LVIS Blue as a low porosity stent and coil adjuvantJournal of NeuroInterventional Surgery, 2018
- A tracking system to calculate patient skin dose in real‐time during neurointerventional procedures using a biplane x‐ray imaging systemMedical Physics, 2016
- Endovascular coil embolization of a very small ruptured aneurysm using a novel microangiographic technique: technical noteJournal of NeuroInterventional Surgery, 2012
- Use of the Microangiographic Fluoroscope for Coiling of Intracranial AneurysmsNeurosurgery, 2011
- A theoretical and experimental evaluation of the microangiographic fluoroscope: A high‐resolution region‐of‐interest x‐ray imagerMedical Physics, 2011
- AAPM/RSNA Physics Tutorial for Residents: Physics of Flat-Panel Fluoroscopy SystemsRadioGraphics, 2011
- The Pipeline Embolization Device for the Intracranial Treatment of Aneurysms TrialAmerican Journal of Neuroradiology, 2010