The effect of various canal contents on the accuracy of two electronic apex locators in detecting different size of root perforations
- 1 June 2021
- journal article
- research article
- Published by Marmara University in Clinical and Experimental Health Sciences
- Vol. 11 (2) , 258-262
Abstract
Objective: This study was to investigate the effect of different canal contents on the accuracy of Gold Reciproc Motor (GRM; VDW, Munich, Germany) and DentaPort ZX (Morita Co, Kyoto, Japan) in the determination of artificial root perforations. Methods: Forty mandibular premolar teeth were included in this study. The crowns of the teeth were removed and the root lengths were standardized to be 14 +/- 1 mm. Roots were divided into 2 groups (n=20). Artificial root perforations were created 0.5 +/- 0.1mm and 1 +/- 0.1 mm in size respectively. The actual lengths (AL) up to the perforation areas were measured under the stereomicroscope. Electronic measurements (EL) were obtained by GRM and DentaPort ZX in dry conditions and the presence of NaOCl (5.25%), EDTA (17%) and blood. The difference was calculated by subtracting the ALs from the ELs. This difference was positive when the measurement was longer than the AL and negative when the measurement was shorter. The Friedman and Wilcoxon signed-rank tests were used to analyze the data (p<0.05). Results: There were no significant differences amongst the different intracanal conditions for both apex locators in teeth with perforation of 1 mm. In the teeth with a perforation of 0.5 mm DentaPort ZX measurements were not affected by intracanal conditions and the most accurate measurement was obtained when the canal was dry with GRM. Conclusion: Within the limitation of this study, intracanal conditions did not affect the measurements of DentaPort ZX in both perforation sizes, whereas in 0.5 mm perforation size, GRM measurements were affected by intracanal conditions..Keywords
This publication has 29 references indexed in Scilit:
- Comparison of two electronic apex locators on human cadaversClinical Oral Investigations, 2015
- Comparison of various current electronic apex locators to determine the working length using the clearing techniqueNigerian Journal of Clinical Practice, 2015
- Accuracy of electronic apex locators to detect root canal perforations with inserted metallic posts: an ex vivo studyHead & Face Medicine, 2014
- Accuracy and Repeatability of 3 Apex Locators in Locating Root Canal Perforations: An Ex Vivo StudyJournal of Endodontics, 2014
- Accuracy and Reliability of Working Length Determination by Gold Reciproc Motor in Reciprocating MovementJournal of Endodontics, 2014
- In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigantsInternational Endodontic Journal, 2012
- The Use of Cone-beam Computed Tomography and Digital Periapical Radiographs to Diagnose Root PerforationsJournal of Endodontics, 2011
- Accuracy of an electronic apex locator in primary teeth with root resorptionInternational Endodontic Journal, 2009
- The influence of irrigating solutions on the accuracy of the electronic apex locator facility in the Tri Auto ZX handpieceInternational Endodontic Journal, 2007
- Reliability of different electronic apex locators to detect root perforations in vitroInternational Endodontic Journal, 1997