Polyethylene insert damage in unicondylar knee replacement: A comparison of in vivo function and in vitro simulation
- 28 December 2009
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 224 (7) , 823-830
Abstract
Modification of knee joint wear simulation methods has included ‘anatomic attachment’ of unicondylar knee replacements (UKR) onto synthetic femurs with material properties and morphology similar to human femurs. The present study assesses the effect of such modification by comparing the damage patterns on UKR polyethylene inserts after in vitro simulation using standard and modified simulation methods with those on inserts retrieved after in vivo function. Three groups of UKR inserts were evaluated after retrieval (Explant Group, n = 17) or after knee joint wear simulation with the components attached to standard metal blocks (Standard Group, n = 6) or synthetic femurs (Anatomic Group, n = 6). All UKR had similar non-conforming articular surfaces. Articular damage patterns (mode, frequency, and area) were quantified using digital image photogrammetry. Although some common damage modes were noted, knee joint wear simulation with standard or ‘anatomic’ attachment did not generate damage pattern sizes similar to the explanted UKR. A focal damage pattern consistent with contact between the metal femoral articular surface and the polyethylene inserts was evident on all inserts, but only the Explant Group had evidence of dispersed damage dominated by abrasive modes. Synthetic femurs added complexity to the wear simulation without generating wear patterns substantially more similar to those observed on retrieved inserts.Keywords
This publication has 34 references indexed in Scilit:
- Comparison of polyethylene tibial insert damage from in vivo function and in vitro wear simulationJournal of Orthopaedic Research, 2009
- Relationship between obesity and early failure of total knee prosthesesBMC Musculoskeletal Disorders, 2009
- Implant fixation in knee replacement: Preliminary in vitro comparison of ceramic and metal cemented femoral componentsThe Knee, 2009
- A New Method of In Vitro Wear Assessment of the UHMWPE Tibial Insert in Total Knee ReplacementArtificial Organs, 2008
- Unicompartmental knee prostheses:in vitrowear assessment of the menisci tibial insert after two different fixation methodsPhysics in Medicine & Biology, 2008
- Stair ascent kinematics affect UHMWPE wear and damage in total knee replacementsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2006
- Retrieval, experimental, and computational assessment of the performance of total knee replacementsJournal of Orthopaedic Research, 2006
- The influence of design, materials and kinematics on the in vitro wear of total knee replacementsJournal of Biomechanics, 2005
- The effect of displacement control input parameters on tibiofemoral prosthetic knee wearWear, 2001
- Retrieval analysis of total knee prostheses: A method and its application to 48 total condylar prosthesesJournal of Biomedical Materials Research, 1983