Does the use of additional X-ray beam filtration during cine acquisition reduce clinical image quality and effective dose in cardiac interventional imaging?
- 20 February 2014
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 162 (4) , 597-604
Abstract
The impact of spectral filtration in digital (‘cine’) acquisition was investigated using a flat panel cardiac interventional X-ray imaging system. A 0.1-mm copper (Cu) and 1.0-mm aluminium (Al) filter added to the standard acquisition mode created the filtered mode for comparison. Image sequences of 35 patients were acquired, a double-blind subjective image quality assessment was completed and dose–area product (DAP) rates were calculated. Entrance surface dose (ESD) and effective dose (E) rates were determined for 20- and 30-cm phantoms. Phantom ESD fell by 28 and 41 % and E by 1 and 0.7 %, for the 20- and 30-cm phantoms, respectively, when using the filtration. Patient DAP rates fell by 43 % with no statistically significant difference in clinical image quality. Adding 0.1-mm Cu and 1.0-mm Al filtration in acquisition substantially reduces patient ESD and DAP, with no significant change in E or clinical image quality.Keywords
This publication has 25 references indexed in Scilit:
- INVESTIGATION OF THE PRACTICAL ASPECTS OF AN ADDITIONAL 0.1 MM COPPER X-RAY SPECTRAL FILTER FOR CINE ACQUISITION MODE IMAGING IN A CLINICAL CARE SETTINGHealth Physics, 2010
- Dose optimization in pediatric cardiac x‐ray imagingMedical Physics, 2010
- Comparison of KAP values for CAG examinations: the influence of parameter settingsRadiation Protection Dosimetry, 2010
- An estimate of the collective dose to the European population from cardiac X-ray proceduresBritish Journal of Radiology, 2008
- A large-scale multicentre study in Belgium of dose area product values and effective doses in interventional cardiology using contemporary X-ray equipmentRadiation Protection Dosimetry, 2007
- Audit of radiation dose to patients during coronary angiographyIndian Journal of Medical Sciences, 2007
- X-Ray Dose Reduction in Fluoroscopically Guided Electrophysiology ProceduresPacing and Clinical Electrophysiology, 2006
- Does Digital Acquisition Reduce Patients' Skin Dose in Cardiac Interventional Procedures? An Experimental StudyAmerican Journal of Roentgenology, 2004
- Radiation Burns as a Severe Complication of Fluoroscopically Guided Cardiological InterventionsJournal of Interventional Cardiology, 2004
- Patient dosimetry during coronary interventions: a comprehensive analysisAmerican Heart Journal, 2004