A dual two dimensional electronic portal imaging device transit dosimetry model based on an empirical quadratic formalism
- 1 July 2015
- journal article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 88 (1051) , 20140645
Abstract
This study describes a two dimensional electronic portal imaging device (EPID) transit dosimetry model that can predict either: (1) in-phantom exit dose, or (2) EPID transit dose, for treatment verification. The model was based on a quadratic equation that relates the reduction in intensity to the equivalent path length (EPL) of the attenuator. In this study, two sets of quadratic equation coefficients were derived from calibration dose planes measured with EPID and ionization chamber in water under reference conditions. With two sets of coefficients, EPL can be calculated from either EPID or treatment planning system (TPS) dose planes. Consequently, either the in-phantom exit dose or the EPID transit dose can be predicted from the EPL. The model was tested with two open, five wedge and seven sliding window prostate and head and neck intensity-modulated radiation therapy (IMRT) fields on phantoms. Results were analysed using absolute gamma analysis (3%/3 mm). The open fields gamma pass rates were >96.8% for all comparisons. For wedge and IMRT fields, comparisons between predicted and TPS-computed in-phantom exit dose resulted in mean gamma pass rate of 97.4% (range, 92.3-100%). As for the comparisons between predicted and measured EPID transit dose, the mean gamma pass rate was 97.5% (range, 92.6-100%). An EPID transit dosimetry model that can predict in-phantom exit dose and EPID transit dose was described and proven to be valid. The described model is practical, generic and flexible to encourage widespread implementation of EPID dosimetry for the improvement of patients' safety in radiotherapy.Keywords
This publication has 42 references indexed in Scilit:
- In aqua vivo EPID dosimetryMedical Physics, 2011
- Implementation of EPID transit dosimetry based on a through-air dosimetry algorithmMedical Physics, 2011
- In vivo dose verification from back projection of a transit dose measurement on the central axis of photon beamsPhysica Medica, 2011
- 3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetryRadiotherapy and Oncology, 2010
- A novel method for patient exit and entrance dose prediction based on water equivalent path length measured with an amorphous silicon electronic portal imaging devicePhysics in Medicine & Biology, 2009
- EPID dosimetry must soon become an essential component of IMRT quality assuranceMedical Physics, 2009
- 3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CTRadiotherapy and Oncology, 2008
- A dose delivery verification method for conventional and intensity‐modulated radiation therapy using measured field fluence distributionsMedical Physics, 2003
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- A 4-MV CT scanner for radiation therapy; spectral properties of the therapy beamMedical Physics, 1983