Patient doses from screen-film and full-field digital mammography in a population-based screening programme
Authors:
- 17 February 2011
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 148 (1) , 65-73
Abstract
The aim of this study was to compare mean glandular dose (MGD) in all full-field digital mammography (FFDM) and screen film mammography (SFM) systems used in a national mammography screening program. MGD from 31 screening units (7 FFDM and 24 SFM), based on an average of 50 women at each screening unit, representing 12 X-ray models (6 FFDM and 6 SFM) from five different manufacturers were calculated. The MGD was significantly lower for FFDM compared with SFM (craniocaudal): 1.19 versus 1.27 mGy, respectively, mediolateral oblique: 1.33 versus 1.45 mGy, respectively), but not all of the FFDM units provided lower doses than the SFM units. Comparing FFDMs, the photon counting scanning-slit technology provides significantly lower MGDs than direct and indirect conversion digital technology. The choice of target/filter combination influences the MGD, and has to be optimised with regard to breast thickness.Keywords
This publication has 17 references indexed in Scilit:
- Optimization of exposure parameters in full field digital mammographyMedical Physics, 2008
- Using the European guidelines to evaluate the Norwegian Breast Cancer Screening ProgramEuropean Journal of Epidemiology, 2007
- Dose comparison between screen/film and full-field digital mammographyEuropean Radiology, 2006
- A comparison of the performance of modern screen-film and digital mammography systemsPhysics in Medicine & Biology, 2005
- A survey of patient dose and clinical factors in a full-field digital mammography systemRadiation Protection Dosimetry, 2005
- Evaluation of detector dynamic range in the x-ray exposure domain in mammography: A comparison between film-screen and flat panel detector systemsMedical Physics, 2003
- Dose reduction in full-field digital mammography: an anthropomorphic breast phantom studyBritish Journal of Radiology, 2003
- Optimization of technique factors for a silicon diode array full‐field digital mammography system and comparison to screen‐film mammography with matched average glandular doseMedical Physics, 2003
- Average glandular dose with amorphous silicon full-field digital mammography - Clinical resultsRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2002
- Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocolPhysics in Medicine & Biology, 2000