Spontaneous rotation of a toric implantable collamer lens related to abnormal ciliary body morphology: a case report
Open Access
- 28 August 2020
- journal article
- case report
- Published by Springer Nature in BMC Ophthalmology
- Vol. 20 (1) , 1-4
Abstract
Background: This is a case of spontaneous toric implantable collamer lens (TICL) rotation that occurred twice in the left eye of a patient. Case presentation: A 24-year-old gentleman received TICL implantation for treatment of myopic astigmatism encountered with spontaneous rotation of approximately 90° from its original position. TICL reposition procedure was performed with visual outcome of 20/16. Surprisingly, a re-rotation of TICL occurred. The patient underwent a replacement of non-toric ICL with a larger size after careful re-evaluation with final visual outcome of 20/20. A short and small ciliary process with shallow ciliary sulcus and posteriorly positioned ciliary body was found by Ultrasound Biomicroscopy (UBM). Conclusions: The unique morphology of the ciliary body may have a potential association with the vault and instability of implanted ICL. Careful examination of the ciliary body morphology is essential in preoperative evaluation.Keywords
Funding Information
- the research project of Science Research Foundation of Aier Eye Hospital Group (AF1909D2)
This publication has 11 references indexed in Scilit:
- The Implantable Collamer Lens with a central port: review of the literatureClinical Ophthalmology, 2018
- Analysis of intraocular positions of posterior implantable collamer lens by full-scale ultrasound biomicroscopyBMC Ophthalmology, 2018
- Quantitative Measurements of the Ciliary Body in Eyes With Acute Primary-Angle ClosureInvestigative Ophthalmology & Visual Science, 2016
- Comparison of Ciliary Body Anatomy between American Caucasians and Ethnic Chinese Using Ultrasound BiomicroscopyCurrent Eye Research, 2015
- Outcomes and possible risk factors associated with axis alignment and rotational stability after implantation of the Toric implantable collamer lens for high myopic astigmatism2012
- Improving accuracy of phakic intraocular lens sizing using high-frequency ultrasound biomicroscopyJournal of Cataract & Refractive Surgery, 2011
- Spontaneous Rotation of a Toric Implantable Collamer LensCase Reports in Ophthalmology, 2010
- Axis Alignment and Rotational Stability After Implantation of the Toric Implantable Collamer Lens for Myopic AstigmatismJournal of Refractive Surgery, 2009
- Postoperative Astigmatism and Axis Stability After Implantation of the STAAR Toric Implantable Collamer LensJournal of Refractive Surgery, 2009
- Simple method for accurate alignment in toric phakic and aphakic intraocular lens implantationJournal of Cataract & Refractive Surgery, 2008