Reliability of Point-of-Care International Normalized Ratio Measurements in Various Patient Populations
Authors:
Open Access
- 1 March 2020
- journal article
- research article
- Published by Wolters Kluwer Health in Point of Care: The Journal of Near-Patient Testing & Technology
- Vol. 19 (1) , 12-18
Abstract
Purpose The aim of this study was to determine if the Coagsense point-of-care (POC) instrument provides more reliable international normalized ratio (INR) measurements than Coagucheck XS POC in comparison to the Stago laboratory instrument in different disease states. Methods This was a prospective study of outpatient warfarin patients comparing venous Stago INR to fingerstick INR on the Coagsense and Coagucheck XS POC meters. Patients were invited to study if they had an of INR 2.0 to 5.0 and had a medical history of antiphospholipid syndrome, hypercoagulable disorder, autoimmune condition, peripheral vascular disease, mechanical heart valve, atrial fibrillation, or deep vein thrombosis/pulmonary embolism/cerebrovascular accident history. Results Seventy-seven patients were enrolled. Coagsense correlated well (92% of INRs within 20% of Stago, 64% of INRs within 0.2 of Stago, overall INR bias of 0.1 or 4%). Six patients had greater than 20% POC INR bias, which could have resulted in 4 warfarin dosing errors. Coagucheck XS INRs correlated poorly (49% within 20% of Stago, 10% of INRs were within 0.2 of Stago, overall INR bias of 0.66 or 25.7%). Forty-one patients had greater than 20% POC INR bias in all diseases, which could have resulted in 28 warfarin dosing errors. The average Coagucheck XS INR bias (0.46–1.3 INR) increased with each 0.5 increase in laboratory INR, whereas Coagsense bias remained stable (0.1–0.25) as INR increased up to 4.3. Two patients correlated well on Coagucheck XS but not Coagsense. Conclusion Coagsense correlated better than Coagucheck XS and did not show increasing bias as INR increased. Both POC instruments had higher INR variability in 4 disease states (antiphospholipid syndrome, autoimmune, peripheral vascular disease, and hypercoagulable). Patient-specific laboratory correlations may be needed on each POC device.Keywords
This publication has 18 references indexed in Scilit:
- Evaluation of a new PT‐INR monitoring system in patients with the antiphospholipid syndromeInternational Journal of Laboratory Hematology, 2016
- Correction of Point-of-Care INR Results in Warfarin PatientsPoint of Care: The Journal of Near-Patient Testing & Technology, 2016
- Validation of the CoaguChek XS international normalised ratio point-of-care analyser in patients at Charlotte Maxeke Johannesburg Academic Hospital, South AfricaSouth African Medical Journal, 2016
- Warfarin Monitoring in Antiphospholipid Syndrome and Lupus AnticoagulantAnnals of Pharmacotherapy, 2014
- Transiently Increased Variation Between a Point-of-Care and Laboratory INR Method After a Long Period of CorrelationAmerican Journal of Clinical Pathology, 2013
- Comparison of International Normalized Ratios provided by two point-of-care devices and laboratory-based venipuncture in a pharmacist-managed anticoagulation clinicAmerican Journal of Health-System Pharmacy, 2010
- Factor X and factor II activity levels do not always agree in warfarin-treated lupus anticoagulant patientsBlood Coagulation & Fibrinolysis, 2010
- Controle do tempo de protrombina em sangue capilar e venoso em pacientes com anticoagulação oral: correlação e concordânciaArquivos Brasileiros de Cardiologia, 2007
- Cardiac involvement in the antiphospholipid syndromeLupus, 2005
- Guidelines for implementation of patient self-testing and patient self-management of oral anticoagulation. International consensus guidelines prepared by International Self-Monitoring Association for Oral AnticoagulationInternational Journal of Cardiology, 2005