Relation between denaturation time measured by optical coherence reflectometry and thermal lesion depth during radiofrequency cardiac ablation: Feasibility numerical study
- 23 November 2017
- journal article
- research article
- Published by Wiley in Lasers in Surgery and Medicine
- Vol. 50 (3) , 222-229
Abstract
Background/Objective: Radiofrequency (RF) catheter ablation is a minimally invasive medical procedure used to thermally destroy the focus of cardiac arrhythmias. Novel optical techniques are now being integrated into RF catheters in order to detect the changes in tissue properties. Loss of birefringence due to fiber denaturation at around 70°C is related to changes in accumulated phase retardation and can be measured by polarization‐sensitive optical coherence reflectometry (PS‐OCR). Since irreversible thermal lesions are produced when the tissue reaches 50°C, our goal was to seek the mathematical relationship between both isotherms.Materials and Methods: A two‐dimensional model based on a coupled electric‐thermal problem was built and solved using the finite element method. The model consisted of cardiac tissue, blood, and a non‐irrigated electrode with a sensor embedded in its tip to maintain a specific target electrode temperature. Computer simulations were conducted by varying the tissue characteristics. Lesion depth was estimated by the 50°C isotherm, while the denaturation time (TD) was taken as the time at which the 70°C isotherm reached a depth of 0.75 mm (which corresponds to the optical depth reached by PS‐OCR technology).Results: A strong correlation (R2 > 0.83) was found between TD and lesion depth and an even stronger correlation (R2 > 0.96) was found between TD and the time required to achieve a specific lesion depth. For instance, the ablation time required to ensure a minimum lesion depth of 3 mm was 1.33 × TD + 3.93 × seconds.Conclusions: The computer results confirmed the strong relationship between denaturation time and lesion depth and suggest that measuring denaturation time by PS‐OCR could provide information on the ablation time required to reach a specific lesion depth. Lasers Surg. Med. 50:222–229, 2018.Keywords
Funding Information
- Ministerio de Economía y Competitividad
- Air Force Office of Scientific Research
This publication has 40 references indexed in Scilit:
- The quest for durable lesions in catheter ablation of atrial fibrillation - technological advances in radiofrequency catheters and balloon devicesExpert Review of Medical Devices, 2017
- Anatomical predisposing factors of transmural thermal injury after pulmonary vein isolationEP Europace, 2017
- Avoiding oesophageal injury during cardiac ablation: insights gained from mediastinal anatomyEP Europace, 2017
- Polarimetric assessment of healthy and radiofrequency ablated porcine myocardial tissueJournal of Biophotonics, 2015
- Three‐Dimensional Optoacoustic Monitoring of Lesion Formation in Real Time During Radiofrequency Catheter AblationJournal of Cardiovascular Electrophysiology, 2014
- Left Atrial Anatomy Relevant to Catheter AblationCardiology Research and Practice, 2014
- Locations of High Contact Force During Left Atrial Mapping in Atrial Fibrillation PatientsCirculation: Arrhythmia and Electrophysiology, 2013
- The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA studyHeart Rhythm, 2012
- In vivo intracardiac optical coherence tomography imaging through percutaneous access: toward image-guided radio-frequency ablationJournal of Biomedical Optics, 2011
- Lesion Dimensions During Temperature-Controlled Radiofrequency Catheter Ablation of Left Ventricular Porcine MyocardiumCirculation, 1999