A self calibrating, magnetic sensor approach accurately positions an aortic damage control stent in a porcine model
Open Access
- 7 December 2023
- journal article
- research article
- Published by BMJ in Trauma Surgery & Acute Care Open
- Vol. 8 (1) , e001220
Abstract
Objectives Non-compressible torso hemorrhage remains a high mortality injury, with difficulty mobilizing resources before exsanguination. Previous studies reported on a retrievable stent graft for damage control and morphometric algorithms for rapid placement, yet fluoroscopy is impractical for the austere environment. We hypothesized that magnetic sensors could be used to position stents relative to an external magnet placed on an anatomic landmark, whereas an electromagnet would allow self-calibration to account for environmental noise. Methods A magnetic sensor alone (MSA) and with integrated stent (MSIS) were examined in a porcine model under anesthesia. A target electromagnet was placed on the xiphoid process (position 0 cm). Sensors were placed in the aorta and measurements obtained at positions 0 cm, +4 cm, and +12 cm from the magnet and compared with fluoroscopy. Sensors were examined under conditions of tachycardia/hypertension, hypotension, vibration, and metal shrapnel to simulate environmental factors that might impact accuracy. General linear models compared mean differences between fluoroscopy and sensor readings. Results Both sensors were compatible with a 10 French catheter system and provided real-time assessment of the distance between the sensor and magnetic target in centimeters. Mean differences between fluoroscopy and both magnetic sensor readings demonstrated accuracy within ±0.5 cm for all but one condition at 0 cm and +4 cm, whereas accuracy decreased at +12 cm from the target. Using the control as a reference, there was no significant difference in mean differences between fluoroscopy and both MSA or MSIS readings at 0 cm and +4 cm for all conditions. The system retained effectiveness if the target was overshot. Conclusion Magnetic sensors achieved the highest accuracy as sensors approached the target. Oscillation of the electromagnet on and off effectively accounts for environmental noise. This approach is promising for rapid and accurate placement of damage control retrievable stent grafts when fluoroscopy is impractical. Level of evidence Not applicable.Keywords
Funding Information
- National Institute of Allergy and Infectious Diseases (T32AI106704)
- National Center for Advancing Translational Sciences (UL1TR002733)
- U.S. Department of War (W81XWH-16-2-0062)
This publication has 17 references indexed in Scilit:
- A retrievable rescue stent graft and radiofrequency positioning for rapid control of noncompressible hemorrhageJournal of Trauma and Acute Care Surgery, 2017
- Fixed-Distance Model for Balloon Placement During Fluoroscopy-Free Resuscitative Endovascular Balloon Occlusion of the Aorta in a Civilian PopulationJAMA Surgery, 2017
- Emerging Endovascular Therapies for Non-Compressible Torso HemorrhageShock, 2016
- The AAST prospective Aortic Occlusion for Resuscitation in Trauma and Acute Care Surgery (AORTA) registryJournal of Trauma and Acute Care Surgery, 2016
- Efficacy of a novel fluoroscopy-free endovascular balloon device with pressure release capabilities in the setting of uncontrolled junctional hemorrhageJournal of Trauma and Acute Care Surgery, 2016
- Implementation of resuscitative endovascular balloon occlusion of the aorta as an alternative to resuscitative thoracotomy for noncompressible truncal hemorrhageJournal of Trauma and Acute Care Surgery, 2015
- Functional Outcome after Resuscitative Endovascular Balloon Occlusion of the Aorta of the Proximal and Distal Thoracic Aorta in a Swine Model of Controlled HemorrhageAnnals of Vascular Surgery, 2015
- A novel fluoroscopy-free, resuscitative endovascular aortic balloon occlusion system in a model of hemorrhagic shockJournal of Trauma and Acute Care Surgery, 2013
- Noncompressible Torso HemorrhageSurgical Clinics of North America, 2012
- Radio frequency identification (RFID) applied to surgical spongesSurgical Endoscopy, 2007