Chondroprotective effects of a polycarbonate‐urethane meniscal implant: histopathological results in a sheep model
- 16 July 2010
- journal article
- knee
- Published by Wiley in Knee Surgery, Sports Traumatology, Arthroscopy
- Vol. 19 (2) , 255-263
Abstract
Purpose: Injury or loss of the meniscus generally leads to degenerative osteoarthritic changes in the knee joint. However, few surgical options exist for meniscal replacement. The goal of this study was to examine the ability of a non‐degradable, anatomically shaped artificial meniscal implant, composed of Kevlar®‐reinforced polycarbonate‐urethane (PCU), to prevent progressive cartilage degeneration following complete meniscectomy.Methods: The artificial meniscus was implanted in the knees of mature female sheep following total medial meniscectomy, and the animals were killed at 3‐ and 6‐months post‐surgery. Macroscopic analysis and semi‐quantitative histological analysis were performed on the cartilage of the operated knee and unoperated contralateral control joint.Results: The PCU implants remained well secured throughout the experimental period and showed no signs of wear or changes in structural or material properties. Histological analysis showed relatively mild cartilage degeneration that was dominated by loss of proteoglycan content and cartilage structure. However, the total osteoarthritis score did not significantly differ between the control and operated knees, and there were no differences in the severity of degenerative changes between 3 and 6 months post‐surgery.Conclusion: Current findings provide preliminary evidence for the ability of an artificial PCU meniscal implant to delay or prevent osteoarthritic changes in knee joint following complete medial meniscectomy.Keywords
This publication has 40 references indexed in Scilit:
- Performance of a Sterile Meniscal Allograft in an Ovine ModelClinical Orthopaedics and Related Research, 2009
- In vitro macrophage response to polyethylene and polycarbonate‐urethane particlesJournal of Biomedical Materials Research Part A, 2009
- In vivo response to cross‐linked polyethylene and polycarbonate‐urethane particlesJournal of Biomedical Materials Research Part A, 2009
- Implantation of a synthetic meniscal scaffold improves joint contact mechanics in a partial meniscectomy cadaver modelJournal of Biomedical Materials Research Part A, 2009
- Tribological behavior of thermoplastic polyurethane elastomersMaterials & Design (1980-2015), 2007
- Polyurethane unicondylar knee prostheses: simulator wear tests and lubrication studiesPhysics in Medicine & Biology, 2006
- The consequences of meniscectomyThe Journal of Bone and Joint Surgery. British volume, 2006
- Retropatellar Chondromalacia Associated with Medial Osteoarthritis after Meniscus InjuryEuropean Surgical Research, 2006
- Risk factors for symptomatic knee osteoarthritis fifteen to twenty‐two years after meniscectomyArthritis & Rheumatism, 2004
- Survival analysis of human meniscal transplantationsThe Journal of Bone and Joint Surgery. British volume, 2002