Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis
Open Access
- 10 October 2024
- journal article
- review article
- Published by JMIR Publications Inc. in Journal of Medical Internet Research
- Vol. 26, e62769
Abstract
Background: Photoplethysmography (PPG) is a technology routinely used in clinical practice to assess blood oxygenation (SpO2) and pulse rate (PR). Skin pigmentation may influence accuracy, leading to health outcomes disparities. Objective: This systematic review and meta-analysis primarily aimed to evaluate the accuracy of PPG-derived SpO2 and PR by skin pigmentation. Secondarily, we aimed to evaluate statistical biases and the clinical relevance of PPG-derived SpO2 and PR according to skin pigmentation. Methods: We identified 23 pulse oximetry studies (n=59,684; 197,353 paired SpO2-arterial blood observations) and 4 wearable PR studies (n=176; 140,771 paired PPG-electrocardiography observations). We evaluated accuracy according to skin pigmentation group by comparing SpO2 accuracy root-mean-square values to the regulatory threshold of 3% and PR 95% limits of agreement values to +5 or –5 beats per minute (bpm), according to the standards of the American National Standards Institute, Association for the Advancement of Medical Instrumentation, and the International Electrotechnical Commission. We evaluated biases and clinical relevance using mean bias and 95% CI. Results: For SpO2, accuracy root-mean-square values were 3.96%, 4.71%, and 4.15%, and pooled mean biases were 0.70% (95% CI 0.17%-1.22%), 0.27% (95% CI –0.64% to 1.19%), and 1.27% (95% CI 0.58%-1.95%) for light, medium, and dark pigmentation, respectively. For PR, 95% limits of agreement values were from –16.02 to 13.54, from –18.62 to 16.84, and from –33.69 to 32.54, and pooled mean biases were –1.24 (95% CI –5.31 to 2.83) bpm, –0.89 (95% CI –3.70 to 1.93) bpm, and –0.57 (95% CI –9.44 to 8.29) bpm for light, medium, and dark pigmentation, respectively. Conclusions: SpO2 and PR measurements may be inaccurate across all skin pigmentation groups, breaching U.S. Food and Drug Administration guidance and industry standard thresholds. Pulse oximeters significantly overestimate SpO2 for both light and dark skin pigmentation, but this overestimation may not be clinically relevant. PRs obtained from wearables exhibit no statistically or clinically significant bias based on skin pigmentation.Keywords
All Related Versions
This publication has 53 references indexed in Scilit:
- Clinical Outcomes Associated With Overestimation of Oxygen Saturation by Pulse Oximetry in Patients Hospitalized With COVID-19JAMA Network Open, 2023
- Evaluating the Accuracy of Pulse Oximetry in Children According to RaceJAMA Pediatrics, 2023
- Prevention of Bias and Discrimination in Clinical Practice AlgorithmsJAMA, 2023
- Clinical Algorithms, Antidiscrimination Laws, and Medical Device RegulationJAMA, 2023
- Racial effects on Masimo pulse oximetry: a laboratory studyJournal of Clinical Monitoring and Computing, 2022
- Assessment of Racial and Ethnic Differences in Oxygen Supplementation Among Patients in the Intensive Care UnitJAMA Internal Medicine, 2022
- Racial and Ethnic Discrepancy in Pulse Oximetry and Delayed Identification of Treatment Eligibility Among Patients With COVID-19JAMA Internal Medicine, 2022
- Association of Race With Pulse Oximetry Accuracy in Hospitalized ChildrenJAMA Network Open, 2022
- Analysis of Discrepancies Between Pulse Oximetry and Arterial Oxygen Saturation Measurements by Race and Ethnicity and Association With Organ Dysfunction and MortalityJAMA Network Open, 2021
- Pilot Study Assessing the Influence of Skin Type on the Heart Rate Measurements Obtained by Photoplethysmography with the Apple WatchJournal of Medical Systems, 2019