Dual‐focal‐spot single‐detector CT: A simulation study to assess cone‐beam artifacts and noise
Authors:
Open Access
- 1 September 2025
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 26 (9) , e70220
Abstract
Background: Ultra‐wide coverage CT (> 128 detector rows) makes it possible to image a heart or brain in a single rotation but are associated with large cone angles, which can severely degrade the image quality. Purpose: This study evaluate the image quality and artifact levels of a dual‐focal‐spot single‐detector (DFSSD) CT geometry designed to achieve 140 mm z‐axis coverage, through a simulation study. Methods: The DFSSD CT system employs two x‐ray focal spots spaced 90 mm apart along the z‐axis and a 100 mm CT detector. It provides 140 mm z‐axis coverage at the iso‐center. A research CT simulation package (CatSim) was employed throughout this study. To compare the image quality, three system geometries were simulated: (A) single focal spot with 40 mm detector (VCT40), (B) single focal spot with 140 mm detector (VCT140), and (C) the DFSSD CT system. A simulated helical body phantom (45 cm x 30 cm x 16 cm) containing various bony structures was used to assess cone‐beam artifacts and noise uniformity across the edge, central, and intermediate slices under different acquisition modes: axial half scan, axial full scan, and helical scan at various helical pitch (0.5, 0.75, 1.0). All images were reconstructed using a cone‐beam filtered back‐projection algorithm with 3D cone‐angle‐dependent pixel‐wise weighting. Results: Under helical scan conditions (pitch 0.5 to 1.0), all geometries demonstrated similar levels of cone‐beam artifacts and noises. Slightly increased artifacts observed on some slices in VCT140 and DFSSD CT, but overall image quality are acceptable. However, under axial half scan condition, the VCT140 geometry exhibited significantly worse artifacts and image uniformity in intermediate and edge slices compared to VCT40, whereas DFSSD CT showed artifact and image uniformity comparable to VCT40. Conclusion: Simulation results indicate that the DFSSD CT geometry can achieve 140 mm z‐axis coverage while maintaining image quality similar to the VCT40 system.Keywords
This publication has 15 references indexed in Scilit:
- Image quality and dose for a multisource cone‐beam CT extremity scannerMedical Physics, 2017
- Multisource inverse-geometry CT. Part I. System concept and developmentMedical Physics, 2016
- Athletic Cardiac Remodeling in US Professional Basketball PlayersJAMA Cardiology, 2016
- 320-detector row CT coronary angiography: effects of heart rate and heart rate variability on image quality, diagnostic accuracy and radiation exposureBritish Journal of Radiology, 2012
- Handling data redundancy in helical cone beam reconstruction with a cone‐angle‐based window function and its asymptotic approximationMedical Physics, 2007
- A three-dimensional-weighted cone beam filtered backprojection (CB-FBP) algorithm for image reconstruction in volumetric CT—helical scanningPhysics in Medicine & Biology, 2006
- A three-dimensional weighted cone beam filtered backprojection (CB-FBP) algorithm for image reconstruction in volumetric CT under a circular source trajectoryPhysics in Medicine & Biology, 2005
- ECG-Gated Cardiac CTAmerican Journal of Roentgenology, 2004
- Multi‐slice helical CT: Scan and reconstructionMedical Physics, 1999
- New perspectives in the assessment of cardiac chamber dimensions during development and adulthoodJACC, 1992