X-ray–induced acoustic computed tomography: 3D X-ray absorption imaging from a single view
Open Access
- 6 December 2024
- journal article
- research article
- Published by American Association for the Advancement of Science (AAAS) in Science Advances
- Vol. 10 (49) , eads1584
Abstract
Computed tomography (CT) scanners are essential for modern imaging but require around 600 projections from various angles. We present x-ray–induced acoustic computed tomography (XACT), a method that uses radiation-induced acoustic waves for three-dimensional (3D) x-ray imaging. These spherical acoustic waves travel through tissue at 1.5 × 10 3 meters per second, much slower than x-rays, allowing ultrasound detectors to capture them and generate 3D images without mechanical scanning. We validate this theory by performing 3D numerical reconstructions of a human breast from a single x-ray projection and experimentally determining 3D structures of objects at different depths. Achieving resolutions of 0.4 millimeters in the XZ plane and 3.5 millimeters in the XY plane at a depth of 16 millimeters, XACT demonstrates the ability to produce 3D images from one x-ray projection, reducing radiation exposure and enabling gantry-free imaging. XACT shows great promise for biomedical and nondestructive testing applications, potentially replacing conventional CT.Keywords
This publication has 67 references indexed in Scilit:
- Photoacoustic Tomography: In Vivo Imaging from Organelles to OrgansScience, 2012
- k-Wave: MATLAB toolbox for the simulation and reconstruction of photoacoustic wave fieldsJournal of Biomedical Optics, 2010
- Three-dimensional structure determination from a single viewNature, 2009
- X-ray computed tomography of the heartBritish Medical Bulletin, 2009
- Radiologic and Nuclear Medicine Studies in the United States and Worldwide: Frequency, Radiation Dose, and Comparison with Other Radiation Sources—1950–2007Radiology, 2009
- CT: History, Technology, and Clinical AspectsSeminars in Ultrasound, CT and MRI, 2005
- Universal back-projection algorithm for photoacoustic computed tomographyPhysical Review E, 2005
- Observation of ultrasonic emission from edges of therapeutic X-ray beamsPhysics in Medicine & Biology, 1991
- X‐ray‐transmission computed tomographyMedical Physics, 1977
- Computerized transverse axial scanning (tomography): Part 1. Description of systemBritish Journal of Radiology, 1973