Effect of Cement, Implant Connection, and Aging on the Retention Strength of Zirconia Crowns to Ti-Base
Authors:
- André Luiz de Melo MorenoAMAndré Luiz de Melo Moreno
- Daniela Micheline dos SantosDSDaniela Micheline dos Santos
- Clovis de Morais Melo NetoCNClovis de Morais Melo Neto
- Renata Vasconcelos MonteiroRMRenata Vasconcelos Monteiro
- Cássia Cunha de LimaCLCássia Cunha de Lima
- Sandro Basso BitencourtSandro Basso Bitencourt
- Nathaly Vilene de Araújo MorenoNMNathaly Vilene de Araújo Moreno
- Wirley Gonçalves AssunçãoWAWirley Gonçalves Assunção
- Marcelo Coelho GoiatoMGMarcelo Coelho Goiato
Preprint
- 25 April 2025
- preprint
- Published by MDPI AG in Preprints.org
Abstract
Purpose: To analyze the influence of the type of cement, aging, and implant connection regarding the Institute of Excellence in Teaching and Research-IETR Institute of Excellence in Teaching and Research-IETR retention capacity of zirconia crowns cemented onto the Titanium base abutment Material and methods: The specimens was divided into four groups: Ti-Base External Hexagon 5,0x4,7x1,0 mm (n=20) or Ti-Base Morse Taper 5,0x4,7x1,5 mm (n=20), temporary or definitive cement, and mechanical aging (with or without). All the samples underwent a tensile test. Results: Data analysis was assessed using the Shapiro-Wilk, three-way ANOVA with repeated measures, and Tukey's test (P < 0.05). The failure mode was also recorded. The Morse taper implants presented a statisti-cally significantly higher average retention strength than the external hexagon implants. Groups III and IV presented the highest value compared to the other groups (565 and 491 N). Most mixed failures were found for all groups. Conclusion: Mechanical aging imped-ed the retention force of milled zirconia crowns cemented on titanium base abutments with resin cement on Morse taper implants had greater retention force compared to exter-nal hexagon implants.Keywords
All Related Versions
This publication has 32 references indexed in Scilit:
- Retention Forces of Monolithic CAD/CAM Crowns Adhesively Cemented to Titanium Base Abutments—Effect of Saliva Contamination Followed by Cleaning of the Titanium Bond SurfaceMaterials, 2021
- Retention of zirconia crowns to Ti-base abutments: effect of luting protocol, abutment treatment and autoclave sterilization.Journal of Prosthodontic Research, 2021
- Three-dimensional misfit between Ti-Base abutments and implants evaluated by replica techniqueJournal of Applied Oral Science, 2020
- A systematic review of the survival and complication rates of zirconia‐ceramic and metal‐ceramic single crownsClinical Oral Implants Research, 2018
- The Effect of CAD/CAM Crown Material and Cement Type on Retention to Implant AbutmentsJournal of Prosthodontics, 2018
- Stress distribution in implant-supported prostheses using different connection systems and cantilever lengths: digital photoelasticityJournal of Medical Engineering & Technology, 2016
- Prosthetic Platforms in Implant DentistryJournal of Craniofacial Surgery, 2011
- Photoelastic Analysis of Cemented or Screwed Implant-Supported Prostheses With Different Prosthetic ConnectionsJournal of Oral Implantology, 2011
- Crown pull-off test (crown retention test) to evaluate the bonding effectiveness of luting agentsDental Materials, 2010
- Fracture resistance of implant-supported screw- versus cement-retained porcelain fused to metal single crowns: SEM fractographic analysisDental Materials, 2007