    Evaluation of the uncertainty of the spectral UV irradiance measured by double- and single-monochromator Brewer spectrophotometers

    by Carmen González, José M. Vilaplana, Alberto Redondas, Javier López-Solano, José M. San Atanasio, Richard Kift, Andrew R. D. Smedley, Pavel Babal, Ana Díaz, Nis Jepsen, et al.

    Brewer instruments are robust, widely used instruments that have been monitoring global solar ultraviolet (UV) irradiance since the 1990s, playing a key role in UV research. Unfortunately, the uncertainties of these measurements are rarely evaluated due to the difficulties involved in characterising the instruments. This evaluation is essential to determine the quality of the measurements as well as their comparability to other datasets. In this study, eight double- and two single-monochromator Brewers are characterised, and the uncertainty of their global UV measurements is estimated using the Monte Carlo method. This methodology is selected because it provides reliable uncertainty estimations and considers the nonlinearity of certain steps in the UV processing algorithm. The combined standard uncertainty depends on the Brewer instrument, varying between 2.5 % and 4 % between 310 and 350 nm. These uncertainties arise primarily from radiometric stability, cosine correction, and the uncertainty of the lamp used during calibration. At shorter wavelengths, the differences between single- and double-monochromator Brewers increase. For example, at 296 nm and a solar zenith angle (SZA) of 40°, the relative uncertainties of single Brewers range between 11 % and 23 %, whereas double Brewers have uncertainties of 3 %–5 %. As the measured wavelength decreases, the correction of stray light (for single Brewers), dark counts, and noise become the dominant sources of uncertainty. These results indicate that the accuracy of fully characterised double Brewers is sufficient for biological studies and trend detection, whereas single Brewers might be limited to wavelengths and SZAs below 305 nm and 70°, respectively.

  • Open Access

    Reply on RC3

    by Carmen González Hernández

    Brewer spectroradiometers are robust, widely used instruments that have been monitoring global solar ultraviolet (UV) irradiance since the 1990s, playing a key role in solar UV research. Unfortunately, the uncertainties of these measurements are rarely evaluated due to the difficulties involved in the uncertainty propagation. This evaluation is essential to determine the quality of the measurements as well as their comparability to other measurements. In this study, eight double- and two single-monochromator Brewers are characterised and the uncertainty of their global UV measurements is estimated using the Monte Carlo method. This methodology is selected as it provides reliable uncertainty estimations and considers the nonlinearity of the UV processing algorithm. The combined standard uncertainty depends on the Brewer, varying between 2.5 % and 4 % for the 300–350 nm region. For wavelengths below 300 nm, the differences between single- and double-monochromator Brewers increase, due to stray light and dark counts. For example, at 295 nm, the relative uncertainties of single Brewers range between 11–14 % while double Brewers have uncertainties of 4–7 %. These uncertainties arise primarily from radiometric stability, the application of cosine correction, and the irradiance of the lamp used during the instrument calibration. As the intensity of the UV irradiance measured decreases, dark counts, stray light (for single Brewers), and noise become the dominant sources of uncertainty. These results indicate that the overall uncertainty of a Brewer spectroradiometer could be greatly reduced by increasing the frequency of radiometric calibration and improving the traditional entrance optics.

  • Open Access

    Reply on RC4

    by Carmen González Hernández

    Brewer spectroradiometers are robust, widely used instruments that have been monitoring global solar ultraviolet (UV) irradiance since the 1990s, playing a key role in solar UV research. Unfortunately, the uncertainties of these measurements are rarely evaluated due to the difficulties involved in the uncertainty propagation. This evaluation is essential to determine the quality of the measurements as well as their comparability to other measurements. In this study, eight double- and two single-monochromator Brewers are characterised and the uncertainty of their global UV measurements is estimated using the Monte Carlo method. This methodology is selected as it provides reliable uncertainty estimations and considers the nonlinearity of the UV processing algorithm. The combined standard uncertainty depends on the Brewer, varying between 2.5 % and 4 % for the 300–350 nm region. For wavelengths below 300 nm, the differences between single- and double-monochromator Brewers increase, due to stray light and dark counts. For example, at 295 nm, the relative uncertainties of single Brewers range between 11–14 % while double Brewers have uncertainties of 4–7 %. These uncertainties arise primarily from radiometric stability, the application of cosine correction, and the irradiance of the lamp used during the instrument calibration. As the intensity of the UV irradiance measured decreases, dark counts, stray light (for single Brewers), and noise become the dominant sources of uncertainty. These results indicate that the overall uncertainty of a Brewer spectroradiometer could be greatly reduced by increasing the frequency of radiometric calibration and improving the traditional entrance optics.

  • Open Access

    Reply on RC2

    by Carmen González Hernández

    Brewer spectroradiometers are robust, widely used instruments that have been monitoring global solar ultraviolet (UV) irradiance since the 1990s, playing a key role in solar UV research. Unfortunately, the uncertainties of these measurements are rarely evaluated due to the difficulties involved in the uncertainty propagation. This evaluation is essential to determine the quality of the measurements as well as their comparability to other measurements. In this study, eight double- and two single-monochromator Brewers are characterised and the uncertainty of their global UV measurements is estimated using the Monte Carlo method. This methodology is selected as it provides reliable uncertainty estimations and considers the nonlinearity of the UV processing algorithm. The combined standard uncertainty depends on the Brewer, varying between 2.5 % and 4 % for the 300–350 nm region. For wavelengths below 300 nm, the differences between single- and double-monochromator Brewers increase, due to stray light and dark counts. For example, at 295 nm, the relative uncertainties of single Brewers range between 11–14 % while double Brewers have uncertainties of 4–7 %. These uncertainties arise primarily from radiometric stability, the application of cosine correction, and the irradiance of the lamp used during the instrument calibration. As the intensity of the UV irradiance measured decreases, dark counts, stray light (for single Brewers), and noise become the dominant sources of uncertainty. These results indicate that the overall uncertainty of a Brewer spectroradiometer could be greatly reduced by increasing the frequency of radiometric calibration and improving the traditional entrance optics.

  • Open Access

    Reply on AC1

    by Carmen González, José Manuel Vilaplana, Alberto Redondas, Javier López-Solano, José María San Atanasio, Richard Kift, Andrew R. D. Smedley, Pavel Babal, Ana Díaz, Nis Jepsen, et al.

    Brewer spectroradiometers are robust, widely used instruments that have been monitoring global solar ultraviolet (UV) irradiance since the 1990s, playing a key role in solar UV research. Unfortunately, the uncertainties of these measurements are rarely evaluated due to the difficulties involved in the uncertainty propagation. This evaluation is essential to determine the quality of the measurements as well as their comparability to other measurements. In this study, eight double- and two single-monochromator Brewers are characterised and the uncertainty of their global UV measurements is estimated using the Monte Carlo method. This methodology is selected as it provides reliable uncertainty estimations and considers the nonlinearity of the UV processing algorithm. The combined standard uncertainty depends on the Brewer, varying between 2.5 % and 4 % for the 300–350 nm region. For wavelengths below 300 nm, the differences between single- and double-monochromator Brewers increase, due to stray light and dark counts. For example, at 295 nm, the relative uncertainties of single Brewers range between 11–14 % while double Brewers have uncertainties of 4–7 %. These uncertainties arise primarily from radiometric stability, the application of cosine correction, and the irradiance of the lamp used during the instrument calibration. As the intensity of the UV irradiance measured decreases, dark counts, stray light (for single Brewers), and noise become the dominant sources of uncertainty. These results indicate that the overall uncertainty of a Brewer spectroradiometer could be greatly reduced by increasing the frequency of radiometric calibration and improving the traditional entrance optics.

  • Open Access

    Comment on egusphere-2025-490

    by Carmen González, José Manuel Vilaplana, Alberto Redondas, Javier López-Solano, José María San Atanasio, Richard Kift, Andrew R. D. Smedley, Pavel Babal, Ana Díaz, Nis Jepsen, et al.

    Brewer spectroradiometers are robust, widely used instruments that have been monitoring global solar ultraviolet (UV) irradiance since the 1990s, playing a key role in solar UV research. Unfortunately, the uncertainties of these measurements are rarely evaluated due to the difficulties involved in the uncertainty propagation. This evaluation is essential to determine the quality of the measurements as well as their comparability to other measurements. In this study, eight double- and two single-monochromator Brewers are characterised and the uncertainty of their global UV measurements is estimated using the Monte Carlo method. This methodology is selected as it provides reliable uncertainty estimations and considers the nonlinearity of the UV processing algorithm. The combined standard uncertainty depends on the Brewer, varying between 2.5 % and 4 % for the 300–350 nm region. For wavelengths below 300 nm, the differences between single- and double-monochromator Brewers increase, due to stray light and dark counts. For example, at 295 nm, the relative uncertainties of single Brewers range between 11–14 % while double Brewers have uncertainties of 4–7 %. These uncertainties arise primarily from radiometric stability, the application of cosine correction, and the irradiance of the lamp used during the instrument calibration. As the intensity of the UV irradiance measured decreases, dark counts, stray light (for single Brewers), and noise become the dominant sources of uncertainty. These results indicate that the overall uncertainty of a Brewer spectroradiometer could be greatly reduced by increasing the frequency of radiometric calibration and improving the traditional entrance optics.

  • Open Access

    Comment on egusphere-2025-490

    by Carmen González, José Manuel Vilaplana, Alberto Redondas, Javier López-Solano, José María San Atanasio, Richard Kift, Andrew R. D. Smedley, Pavel Babal, Ana Díaz, Nis Jepsen, et al.

    Brewer spectroradiometers are robust, widely used instruments that have been monitoring global solar ultraviolet (UV) irradiance since the 1990s, playing a key role in solar UV research. Unfortunately, the uncertainties of these measurements are rarely evaluated due to the difficulties involved in the uncertainty propagation. This evaluation is essential to determine the quality of the measurements as well as their comparability to other measurements. In this study, eight double- and two single-monochromator Brewers are characterised and the uncertainty of their global UV measurements is estimated using the Monte Carlo method. This methodology is selected as it provides reliable uncertainty estimations and considers the nonlinearity of the UV processing algorithm. The combined standard uncertainty depends on the Brewer, varying between 2.5 % and 4 % for the 300–350 nm region. For wavelengths below 300 nm, the differences between single- and double-monochromator Brewers increase, due to stray light and dark counts. For example, at 295 nm, the relative uncertainties of single Brewers range between 11–14 % while double Brewers have uncertainties of 4–7 %. These uncertainties arise primarily from radiometric stability, the application of cosine correction, and the irradiance of the lamp used during the instrument calibration. As the intensity of the UV irradiance measured decreases, dark counts, stray light (for single Brewers), and noise become the dominant sources of uncertainty. These results indicate that the overall uncertainty of a Brewer spectroradiometer could be greatly reduced by increasing the frequency of radiometric calibration and improving the traditional entrance optics.

  • Open Access

    Reply on RC1

    by Carmen González Hernández

    Brewer spectroradiometers are robust, widely used instruments that have been monitoring global solar ultraviolet (UV) irradiance since the 1990s, playing a key role in solar UV research. Unfortunately, the uncertainties of these measurements are rarely evaluated due to the difficulties involved in the uncertainty propagation. This evaluation is essential to determine the quality of the measurements as well as their comparability to other measurements. In this study, eight double- and two single-monochromator Brewers are characterised and the uncertainty of their global UV measurements is estimated using the Monte Carlo method. This methodology is selected as it provides reliable uncertainty estimations and considers the nonlinearity of the UV processing algorithm. The combined standard uncertainty depends on the Brewer, varying between 2.5 % and 4 % for the 300–350 nm region. For wavelengths below 300 nm, the differences between single- and double-monochromator Brewers increase, due to stray light and dark counts. For example, at 295 nm, the relative uncertainties of single Brewers range between 11–14 % while double Brewers have uncertainties of 4–7 %. These uncertainties arise primarily from radiometric stability, the application of cosine correction, and the irradiance of the lamp used during the instrument calibration. As the intensity of the UV irradiance measured decreases, dark counts, stray light (for single Brewers), and noise become the dominant sources of uncertainty. These results indicate that the overall uncertainty of a Brewer spectroradiometer could be greatly reduced by increasing the frequency of radiometric calibration and improving the traditional entrance optics.

  • Open Access

    Comment on egusphere-2025-490

    by Carmen González, José Manuel Vilaplana, Alberto Redondas, Javier López-Solano, José María San Atanasio, Richard Kift, Andrew R. D. Smedley, Pavel Babal, Ana Díaz, Nis Jepsen, et al.

    Brewer spectroradiometers are robust, widely used instruments that have been monitoring global solar ultraviolet (UV) irradiance since the 1990s, playing a key role in solar UV research. Unfortunately, the uncertainties of these measurements are rarely evaluated due to the difficulties involved in the uncertainty propagation. This evaluation is essential to determine the quality of the measurements as well as their comparability to other measurements. In this study, eight double- and two single-monochromator Brewers are characterised and the uncertainty of their global UV measurements is estimated using the Monte Carlo method. This methodology is selected as it provides reliable uncertainty estimations and considers the nonlinearity of the UV processing algorithm. The combined standard uncertainty depends on the Brewer, varying between 2.5 % and 4 % for the 300–350 nm region. For wavelengths below 300 nm, the differences between single- and double-monochromator Brewers increase, due to stray light and dark counts. For example, at 295 nm, the relative uncertainties of single Brewers range between 11–14 % while double Brewers have uncertainties of 4–7 %. These uncertainties arise primarily from radiometric stability, the application of cosine correction, and the irradiance of the lamp used during the instrument calibration. As the intensity of the UV irradiance measured decreases, dark counts, stray light (for single Brewers), and noise become the dominant sources of uncertainty. These results indicate that the overall uncertainty of a Brewer spectroradiometer could be greatly reduced by increasing the frequency of radiometric calibration and improving the traditional entrance optics.

  • Open Access

    Evaluation of the uncertainty of the spectral UV irradiance measured by double- and single-monochromator Brewer spectroradiometers

    by Carmen González, José Manuel Vilaplana, Alberto Redondas, Javier López-Solano, José María San Atanasio, Richard Kift, Andrew R. D. Smedley, Pavel Babal, Ana Díaz, Nis Jepsen, et al.

    Brewer spectroradiometers are robust, widely used instruments that have been monitoring global solar ultraviolet (UV) irradiance since the 1990s, playing a key role in solar UV research. Unfortunately, the uncertainties of these measurements are rarely evaluated due to the difficulties involved in the uncertainty propagation. This evaluation is essential to determine the quality of the measurements as well as their comparability to other measurements. In this study, eight double- and two single-monochromator Brewers are characterised and the uncertainty of their global UV measurements is estimated using the Monte Carlo method. This methodology is selected as it provides reliable uncertainty estimations and considers the nonlinearity of the UV processing algorithm. The combined standard uncertainty depends on the Brewer, varying between 2.5 % and 4 % for the 300–350 nm region. For wavelengths below 300 nm, the differences between single- and double-monochromator Brewers increase, due to stray light and dark counts. For example, at 295 nm, the relative uncertainties of single Brewers range between 11–14 % while double Brewers have uncertainties of 4–7 %. These uncertainties arise primarily from radiometric stability, the application of cosine correction, and the irradiance of the lamp used during the instrument calibration. As the intensity of the UV irradiance measured decreases, dark counts, stray light (for single Brewers), and noise become the dominant sources of uncertainty. These results indicate that the overall uncertainty of a Brewer spectroradiometer could be greatly reduced by increasing the frequency of radiometric calibration and improving the traditional entrance optics.