Vivek Maradia
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Research Subjects
Vivek Maradia has a total of 38 co-authors. Here are the top co-authors by number of shared publications:
David Meer
Paul Scherrer Institute, Switzerland
10 shared publicationsSerena Psoroulas
Paul Scherrer Institute, Switzerland
8 shared publicationsDamien Charles Weber
Paul Scherrer Institute
6 shared publicationsAntony John Lomax
Paul Scherrer Institute
5 shared publications- Serena Psoroulas
Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Zurich, Switzerland
5 shared publications Antony J. Lomax
Paul Scherrer Institute, Switzerland; ETH Zurich, Switzerland
4 shared publications
Latest Publications
- Open Access
High-speed proton therapy within a short breath-hold
Proton therapy provides superior dose conformity compared with photon radiotherapy, concentrating radiation within the tumor while sparing adjacent healthy tissue. This advantage has been most effectively realized for static tumors in anatomically stable regions, such as the head and neck. For thoracic and abdominal sites, however, physiological motion remains a critical challenge: because the proton dose distribution is highly sensitive to density variations, long delivery times relative to respiratory motion can compromise accuracy. Existing strategies to accelerate delivery often require substantial hardware modifications or are difficult to translate into routine practice. Here we report an optimization that enables high-speed proton delivery (5 to 10 sec per field) on a commercial synchrocyclotron platform without hardware changes. The method combines high-energy shoot-through beams with Bragg-peak delivery, an optimized nearest-neighbor scanning sequence, and a two-pulse dose regulation scheme. Applied to eight lung cancer cases (target volumes 100 to 1000 cc), the approach achieved full field delivery in under 10 sec compatible with a short breath hold while preserving conformity, dose accuracy, and sparing of organs at risk. This framework provides a practical route to motion robust proton therapy, improving precision, efficiency and patient tolerance. More broadly, it opens a pathway toward widespread clinical adoption of high-speed proton delivery for moving tumors.
Open AccessHigh-speed proton therapy within a short breath-hold
Proton therapy provides superior dose conformity compared with photon radiotherapy, concentrating radiation within the tumor while sparing adjacent healthy tissue. This advantage has been most effectively realized for static tumors in anatomically stable regions, such as the head and neck. For thoracic and abdominal sites, however, physiological motion remains a critical challenge: because the proton dose distribution is highly sensitive to density variations, long delivery times relative to respiratory motion can compromise accuracy. Existing strategies to accelerate delivery often require substantial hardware modifications or are difficult to translate into routine practice. Here we report an optimization that enables high-speed proton delivery (5 to 10 sec per field) on a commercial synchrocyclotron platform without hardware changes. The method combines high-energy shoot-through beams with Bragg-peak delivery, an optimized nearest-neighbor scanning sequence, and a two-pulse dose regulation scheme. Applied to eight lung cancer cases (target volumes 100 to 1000 cc), the approach achieved full field delivery in under 10 sec compatible with a short breath hold while preserving conformity, dose accuracy, and sparing of organs at risk. This framework provides a practical route to motion robust proton therapy, improving precision, efficiency and patient tolerance. More broadly, it opens a pathway toward widespread clinical adoption of high-speed proton delivery for moving tumors.
- 3 September 2025
- Published by Springer Nature in Nature Physics
- p. 1-11
Accelerator technologies for proton and ion beam therapy
- 3 September 2025
- Published by Springer Nature in Nature Physics
- p. 1-11
Accelerator technologies for proton and ion beam therapy
- Open Access
- 29 May 2024
- Published by Springer Nature in Nature Physics
- Vol. 20 (6) , 1033
Author Correction: Demonstration of momentum cooling to enhance the potential of cancer treatment with proton therapy
- 29 May 2024
- Published by Springer Nature in Nature Physics
- Vol. 20 (6) , 1033
Open AccessAuthor Correction: Demonstration of momentum cooling to enhance the potential of cancer treatment with proton therapy
- Open Access
- 3 July 2023
- Published by Springer Nature in Nature Physics
- Vol. 19 (10) , 1437-1444
Demonstration of momentum cooling to enhance the potential of cancer treatment with proton therapy
In recent years, there has been a considerable push towards ultrahigh dose rates in proton therapy to effectively utilize motion mitigation strategies and potentially increase the sparing of healthy tissue through the so-called FLASH effect. However, in cyclotron-based proton therapy facilities, it is difficult to reach ultrahigh dose rates for low-energy beams. The main reason for this lies in the large momentum spread that such beams have after reducing their energy to levels required for proton therapy, incurring large losses in conventionally used momentum or energy selection slits. Here we propose momentum cooling by using a wedge in the energy selection system (instead of a slit) to reduce the momentum spread of the beam without introducing substantial beam losses. We demonstrate this concept in our eye treatment beamline and obtain a factor of two higher transmission, which could eventually halve the treatment delivery time. Furthermore, we show that with a gantry design incorporating this feature, we can achieve almost a factor of 100 higher transmission for a 70 MeV beam compared with conventional cyclotron-based facilities. This concept could enhance the potential of proton therapy by opening up possibilities of treating new indications and reducing the cost.
- 3 July 2023
- Published by Springer Nature in Nature Physics
- Vol. 19 (10) , 1437-1444
Open AccessDemonstration of momentum cooling to enhance the potential of cancer treatment with proton therapy
In recent years, there has been a considerable push towards ultrahigh dose rates in proton therapy to effectively utilize motion mitigation strategies and potentially increase the sparing of healthy tissue through the so-called FLASH effect. However, in cyclotron-based proton therapy facilities, it is difficult to reach ultrahigh dose rates for low-energy beams. The main reason for this lies in the large momentum spread that such beams have after reducing their energy to levels required for proton therapy, incurring large losses in conventionally used momentum or energy selection slits. Here we propose momentum cooling by using a wedge in the energy selection system (instead of a slit) to reduce the momentum spread of the beam without introducing substantial beam losses. We demonstrate this concept in our eye treatment beamline and obtain a factor of two higher transmission, which could eventually halve the treatment delivery time. Furthermore, we show that with a gantry design incorporating this feature, we can achieve almost a factor of 100 higher transmission for a 70 MeV beam compared with conventional cyclotron-based facilities. This concept could enhance the potential of proton therapy by opening up possibilities of treating new indications and reducing the cost.
- Open Access
- 1 January 2023
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 2420 (1) , 012106
A novel intensity compensation method to achieve energy independent beam intensity at the patient location for cyclotron based proton therapy facilities
In cyclotron-based proton therapy facilities, an energy selection system is typically used to lower the beam energy from the fixed value provided by the accelerator (250/230 MeV) to the one needed for the treatment (230-70 MeV). Such a system has the drawback of increase beam emittance and introducing an energy-dependent beam current at the patient location, resulting in energy dependent beam intensity ratios of about 103 between high and low energies. This complicates treatment delivery and challenges patient safety systems. As such, we propose the use of a dual energy degrader method that can reduce beam intensity for high-energy beams. The first degrader is made of high Z material and the second is made of low Z material and are placed next to each other. For high energies (190-230 MeV), we use only the first degrader to increase beam emittance after the degrader and thus lose intensity in the emittance selection collimators. For intermediate energy beams (110-190 MeV) we use the combination of both degraders, whereas for low energy beams (70-110 MeV), only the second degrader limits the increase in emittance. With this approach, energy-independent beam intensities at the patient location can be achieved, whilst localizing beam losses around the degrader.
- 1 January 2023
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 2420 (1) , 012106
Open AccessA novel intensity compensation method to achieve energy independent beam intensity at the patient location for cyclotron based proton therapy facilities
In cyclotron-based proton therapy facilities, an energy selection system is typically used to lower the beam energy from the fixed value provided by the accelerator (250/230 MeV) to the one needed for the treatment (230-70 MeV). Such a system has the drawback of increase beam emittance and introducing an energy-dependent beam current at the patient location, resulting in energy dependent beam intensity ratios of about 103 between high and low energies. This complicates treatment delivery and challenges patient safety systems. As such, we propose the use of a dual energy degrader method that can reduce beam intensity for high-energy beams. The first degrader is made of high Z material and the second is made of low Z material and are placed next to each other. For high energies (190-230 MeV), we use only the first degrader to increase beam emittance after the degrader and thus lose intensity in the emittance selection collimators. For intermediate energy beams (110-190 MeV) we use the combination of both degraders, whereas for low energy beams (70-110 MeV), only the second degrader limits the increase in emittance. With this approach, energy-independent beam intensities at the patient location can be achieved, whilst localizing beam losses around the degrader.
- Open Access
- 1 January 2023
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 2420 (1) , 012107
A novel method of emittance matching to increase beam transmission for cyclotron-based proton therapy facilities: simulation study
In proton therapy, high dose rates can reduce treatment delivery times, allowing for efficient mitigation of tumor motion and increased patient throughput. With cyclotrons however, high dose rates are difficult to achieve for low-energies as, typically, the emittance after the degrader is matched in both transversal planes using circular collimators, which does not provide an optimal matching to the acceptance of the following beamline. Transmission can however be substantially improved by transporting maximum acceptable emittances in both orthogonal planes, but at the cost of gantry angle-dependent beam shapes at isocenter. Here we demonstrate that equal emittances in both planes can be recovered at the gantry entrance using a thin scattering foil, thus ensuring gantry angle-independent beam shapes at the isocenter. We demonstrate in simulation that low-energy beam transmission can be increased by a factor of 3 using this approach compared to the currently used beam optics, whilst gantry angle-independent beam shapes are preserved. We expect that this universal approach could also bring a similar transmission improvement in other cyclotron-based proton therapy facilities.
- 1 January 2023
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 2420 (1) , 012107
Open AccessA novel method of emittance matching to increase beam transmission for cyclotron-based proton therapy facilities: simulation study
In proton therapy, high dose rates can reduce treatment delivery times, allowing for efficient mitigation of tumor motion and increased patient throughput. With cyclotrons however, high dose rates are difficult to achieve for low-energies as, typically, the emittance after the degrader is matched in both transversal planes using circular collimators, which does not provide an optimal matching to the acceptance of the following beamline. Transmission can however be substantially improved by transporting maximum acceptable emittances in both orthogonal planes, but at the cost of gantry angle-dependent beam shapes at isocenter. Here we demonstrate that equal emittances in both planes can be recovered at the gantry entrance using a thin scattering foil, thus ensuring gantry angle-independent beam shapes at the isocenter. We demonstrate in simulation that low-energy beam transmission can be increased by a factor of 3 using this approach compared to the currently used beam optics, whilst gantry angle-independent beam shapes are preserved. We expect that this universal approach could also bring a similar transmission improvement in other cyclotron-based proton therapy facilities.
- Open Access
- 7 November 2022
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 67 (22) , 225005
Universal and dynamic ridge filter for pencil beam scanning particle therapy: a novel concept for ultra-fast treatment delivery
Objective. In pencil beam scanning particle therapy, a short treatment delivery time is paramount for the efficient treatment of moving targets with motion mitigation techniques (such as breath-hold, rescanning, and gating). Energy and spot position change time are limiting factors in reducing treatment time. In this study, we designed a universal and dynamic energy modulator (ridge filter, RF) to broaden the Bragg peak, to reduce the number of energies and spots required to cover the target volume, thus lowering the treatment time. Approach . Our RF unit comprises two identical RFs placed just before the isocenter. Both RFs move relative to each other, changing the Bragg peak’s characteristics dynamically. We simulated different Bragg peak shapes with the RF in Monte Carlo simulation code (TOPAS) and validated them experimentally. We then delivered single-field plans with 1 Gy/fraction to different geometrical targets in water, to measure the dose delivery time using the RF and compare it with the clinical settings. Main results. Aligning the RFs in different positions produces different broadening in the Bragg peak; we achieved a maximum broadening of 2.5 cm. With RF we reduced the number of energies in a field by more than 60%, and the dose delivery time by 50%, for all geometrical targets investigated, without compromising the dose distribution transverse and distal fall-off. Significance . Our novel universal and dynamic RF allows for the adaptation of the Bragg peak broadening for a spot and/or energy layer based on the requirement of dose shaping in the target volume. It significantly reduces the number of energy layers and spots to cover the target volume, and thus the treatment time. This RF design is ideal for ultra-fast treatment delivery within a single breath-hold (5–10 s), efficient delivery of motion mitigation techniques, and small animal irradiation with ultra-high dose rates (FLASH).
- 7 November 2022
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 67 (22) , 225005
Open AccessUniversal and dynamic ridge filter for pencil beam scanning particle therapy: a novel concept for ultra-fast treatment delivery
Objective. In pencil beam scanning particle therapy, a short treatment delivery time is paramount for the efficient treatment of moving targets with motion mitigation techniques (such as breath-hold, rescanning, and gating). Energy and spot position change time are limiting factors in reducing treatment time. In this study, we designed a universal and dynamic energy modulator (ridge filter, RF) to broaden the Bragg peak, to reduce the number of energies and spots required to cover the target volume, thus lowering the treatment time. Approach . Our RF unit comprises two identical RFs placed just before the isocenter. Both RFs move relative to each other, changing the Bragg peak’s characteristics dynamically. We simulated different Bragg peak shapes with the RF in Monte Carlo simulation code (TOPAS) and validated them experimentally. We then delivered single-field plans with 1 Gy/fraction to different geometrical targets in water, to measure the dose delivery time using the RF and compare it with the clinical settings. Main results. Aligning the RFs in different positions produces different broadening in the Bragg peak; we achieved a maximum broadening of 2.5 cm. With RF we reduced the number of energies in a field by more than 60%, and the dose delivery time by 50%, for all geometrical targets investigated, without compromising the dose distribution transverse and distal fall-off. Significance . Our novel universal and dynamic RF allows for the adaptation of the Bragg peak broadening for a spot and/or energy layer based on the requirement of dose shaping in the target volume. It significantly reduces the number of energy layers and spots to cover the target volume, and thus the treatment time. This RF design is ideal for ultra-fast treatment delivery within a single breath-hold (5–10 s), efficient delivery of motion mitigation techniques, and small animal irradiation with ultra-high dose rates (FLASH).
- Open Access
Universal and dynamic ridge filter for pencil beam scanning particle therapy: novel concept for ultra-fast treatment delivery
Purpose In PBS particle therapy, short treatment delivery time is paramount for the efficient treatment of moving targets with motion mitigation techniques (such as breath-hold, rescanning, and gating). Energy and spot position change time are limiting factors in reducing treatment time. In this study, we designed a universal and dynamic energy modulator (ridge filter, RF) to broaden the Bragg peak, to reduce the number of energies and spots required to cover the target volume, thus lowering the treatment time. Methods Our RF unit comprises two identical RFs placed just before the isocenter. Both RFs move relative to each other, changing the Bragg peaks characteristics dynamically. We simulated different Bragg peak shapes with the RF in TOPAS and validated them experimentally. We then delivered single-field plans with 1Gy/fraction to different geometrical targets in water, to measure the dose delivery time using the RF and compare it with the clinical settings. Results Aligning the RFs in different positions produces different broadening in the Bragg peak; we achieved a maximum broadening of 2 cm. With RF we reduced the number of energies in a field by more than 60%, and the dose delivery time by 50%, for all geometrical targets investigated, without compromising the dose distribution transverse and distal fall-off. Conclusions Our novel universal and dynamic RF allows for the adaptation of the Bragg peak broadening for a spot and/or energy layer based on the requirement of dose shaping in the target volume. It significantly reduces the number of energy layers and spots to cover the target volume, and thus the treatment time. This RF design is ideal for ultra-fast treatment delivery within a single breath-hold (5-10 sec), efficient delivery of motion mitigation techniques, and small animal irradiation with ultra-high dose rates (FLASH).
Open AccessUniversal and dynamic ridge filter for pencil beam scanning particle therapy: novel concept for ultra-fast treatment delivery
Purpose In PBS particle therapy, short treatment delivery time is paramount for the efficient treatment of moving targets with motion mitigation techniques (such as breath-hold, rescanning, and gating). Energy and spot position change time are limiting factors in reducing treatment time. In this study, we designed a universal and dynamic energy modulator (ridge filter, RF) to broaden the Bragg peak, to reduce the number of energies and spots required to cover the target volume, thus lowering the treatment time. Methods Our RF unit comprises two identical RFs placed just before the isocenter. Both RFs move relative to each other, changing the Bragg peaks characteristics dynamically. We simulated different Bragg peak shapes with the RF in TOPAS and validated them experimentally. We then delivered single-field plans with 1Gy/fraction to different geometrical targets in water, to measure the dose delivery time using the RF and compare it with the clinical settings. Results Aligning the RFs in different positions produces different broadening in the Bragg peak; we achieved a maximum broadening of 2 cm. With RF we reduced the number of energies in a field by more than 60%, and the dose delivery time by 50%, for all geometrical targets investigated, without compromising the dose distribution transverse and distal fall-off. Conclusions Our novel universal and dynamic RF allows for the adaptation of the Bragg peak broadening for a spot and/or energy layer based on the requirement of dose shaping in the target volume. It significantly reduces the number of energy layers and spots to cover the target volume, and thus the treatment time. This RF design is ideal for ultra-fast treatment delivery within a single breath-hold (5-10 sec), efficient delivery of motion mitigation techniques, and small animal irradiation with ultra-high dose rates (FLASH).
- Open Access
- 11 August 2022
- Published by Frontiers Media SA in Frontiers in Physics
- Vol. 10
Application of a scattering foil to increase beam transmission for cyclotron based proton therapy facilities
In proton therapy, high dose rates can reduce treatment delivery times, allowing for efficient mitigation of tumor motion, as well as increased treatment efficiency and patient throughput. In cyclotron-based facilities, however, high dose rates are difficult to achieve at low-energies. In current facilities, the emittance after the degrader is matched in both transversal planes using circular collimators; this does not provide an optimal matching to the acceptance of the following beamline. However, transmission can be substantially improved by transporting maximum acceptable emittances in the X and Y planes, but at the cost of an elliptical beam shape at the gantry entrance, leading to gantry angle-dependent beam shapes at the isocenter. Here we demonstrate that equal emittances in both planes can be recovered at the gantry entrance using a thin scattering foil, thus ensuring gantry angle-independent beam shape at the isocenter. Using modified beam optics and thin scattering foil placed in the beamline, we demonstrate experimentally that low-energy beam transmission can be increased by a factor of three compared to the currently used beam optics, whilst preserving gantry angle-independent beam shapes, at the cost of a large beam size. We expect that this approach could also bring a similar transmission improvement in other cyclotron-based proton therapy facilities.
- 11 August 2022
- Published by Frontiers Media SA in Frontiers in Physics
- Vol. 10
Open AccessApplication of a scattering foil to increase beam transmission for cyclotron based proton therapy facilities
In proton therapy, high dose rates can reduce treatment delivery times, allowing for efficient mitigation of tumor motion, as well as increased treatment efficiency and patient throughput. In cyclotron-based facilities, however, high dose rates are difficult to achieve at low-energies. In current facilities, the emittance after the degrader is matched in both transversal planes using circular collimators; this does not provide an optimal matching to the acceptance of the following beamline. However, transmission can be substantially improved by transporting maximum acceptable emittances in the X and Y planes, but at the cost of an elliptical beam shape at the gantry entrance, leading to gantry angle-dependent beam shapes at the isocenter. Here we demonstrate that equal emittances in both planes can be recovered at the gantry entrance using a thin scattering foil, thus ensuring gantry angle-independent beam shape at the isocenter. Using modified beam optics and thin scattering foil placed in the beamline, we demonstrate experimentally that low-energy beam transmission can be increased by a factor of three compared to the currently used beam optics, whilst preserving gantry angle-independent beam shapes, at the cost of a large beam size. We expect that this approach could also bring a similar transmission improvement in other cyclotron-based proton therapy facilities.