LZLu Zhang

Lu Zhang

*This is an algorithmically generated author record based on publicly available information.

Current Organizations / Institutes

The organizations/institutes from the scholar's latest article


Lu Zhang has a total of 16 co-authors. Here are the top co-authors by number of shared publications:

  • Gerd Uwe Auffarth

    International Vision Correction Research Centre (IVCRC), Department of Ophthalmology, University of Heidelberg, INF 400, 69120 Heidelberg, Germany

    5 shared publications
  • Sonja Schickhardt

    The David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University Hospital of Heidelberg, Heidelberg, Germany

    4 shared publications
  • Patrick Merz

    David J. Apple International Laboratory for Ocular Pathology and International Vision Correction Research Centre (IVCRC), Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany

    2 shared publications
  • Donald Bs Munro

    The David J Apple International Laboratory for Ocular Pathology, Heidelberg, Germany; and

    1 shared publication
  • Donald John Munro

    The David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University Hospital of Heidelberg, Heidelberg, Germany

    1 shared publication
  • Florian Auerbach

    David J Apple Center for Vision Research, Department of Ophthalmology, University Hospital Heidelberg, Im Neuenheimer Feld 400, 69120 Heidelberg, Germany

    1 shared publication

Latest Publications

  • Open Access

    Evaluation of Parameters and Nozzle Tip Damage after Clinical Use of Three Hydrophilic Intraocular Lens Injector Models

    by Lu Zhang, Sonja Schickhardt, Patrick Merz, Gerd Uwe Auffarth

    Purpose. To assess the nozzle tip damage and the parameters of three different hydrophilic intraocular lens (IOL) injector models. Methods. After routine cataract surgeries at the University Eye Hospital Heidelberg, all the used IOL injectors were collected from the operating room and sent to our laboratory. Nozzle tip damage was assessed under a microscope and graded as follows: no damage (grade 0), slight scratches (1), deep scratches (2), extensions (3), cracks (4), and bursts (5). Each damage grade was assigned a score from 0 to 5, and the total damage score for each injector system was calculated and compared. Nozzle tip parameters (diameters and areas), plunger tip parameters, and tip angles were also measured in each model. Results. The damage scores were (median, Q3-Q1): 1 (1-1) for Accuject, 1 (1-1) for Bluemixs, and 1 (1-1) for RayOne. There was no statistically significant difference in the damage scores between the study groups ( ). The outer cross-sectional vertical and horizontal diameters were 1.69 and 1.69 mm for Accuject, 1.69 and 1.69 mm for Bluemixs, and 1.70 and 1.71 mm for RayOne. Plunger tip areas were 0.78 mm2 for Accjuect, 0.74 mm2 for Bluemixs, and 0.43 mm2 for RayOne. Plunger tip area/inner cross-sectional area of the nozzle tip (%) was 31.2% for RayOne, 66.7% for Accuject, and 63.8% for Bluemixs. The tip angles for three injector models were 56° (Accuject), 56° (Bluemixs), and 44° (RayOne). Conclusions. All the injector models showed mild to moderate damage to the nozzle tip after IOL implantation, even with smaller diameter tips. RayOne resulted in the lowest ratio between plunger tip area and inner cross-sectional area of the nozzle tip and a better distribution of damage categories than the other two groups. All three injector models had relatively small tip parameters. If smaller incisions are required in certain patients, smaller tip parameters should be considered.


  • Open Access

    In-vitro assessment of a novel intraocular lens made of crosslinked polyisobutylene

    by Sonja K. Schickhardt, Grzegorz Łabuz, Donald J. Munro, Ingo Lieberwirth, Lu Zhang, Hui Fang, Gerd U. Auffarth


  • Open Access

    Capsule Dynamics, Implantation, and Explantation of the Smaller Incision, New-Generation Implantable Miniature Telescope: a Miyake–Apple Study

    by Maximilian Cand. Med. Hammer, Sonja Schickhardt, Lu Zhang, Donald Bs Munro, Timur M. Yildirim, Gerd U. Auffarth

    Purpose: The smaller incision, new-generation implantable miniature telescope is a novel implant to optimize vision in retinal patients with central vision loss. Using Miyake–Apple techniques we visualized the device implantation, repositioning, and explantation, while noting capsular bag dynamics. Methods: Using the Miyake–Apple technique, we assessed capsular bag deformation after successful implantation of the device in human autopsy eyes. We assessed rescue strategies for converting a sulcus implantation to a capsular implantation and explantation strategies. We noted the occurrence of posterior capsule striae, zonular stress, and the haptics' arc of contact with the capsular bag after implantation. Results: Acceptable zonular stress was observed during the successful implantation of the SING IMT. When it was implanted in the sulcus, one could reposition the haptics into the bag with two spatulas using counter-pressure in an effective strategy despite inducing tolerable, medium zonular stress. A similar technique, in reverse, allows safe explantation without damaging the rhexis or the bag, while inducing similar medium, tolerable zonular stress. In all eyes we examined, the implant considerably stretches the bag, inducing a capsular bag deformation and posterior capsule striae. Conclusion: The SING IMT can be safely implanted without significant zonular stress. In sulcus implantation and explantation, repositioning of the haptic is achievable without perturbing zonular stress using the presented approaches. It stretches average-sized capsular bags to support its weight. This is achieved by an increased arc of contact of the haptics with the capsular equator.


  • Open Access

    Evaluation of Nozzle Tip Damage in Intraocular Lens Injectors with V-Shaped Notch

    by Lu Zhang, Gerd Auffarth, Sonja Schickhardt, Patrick Merz, Tamer Tandogan

    Introduction: Damage to the nozzle tips of intraocular lens (IOL) injectors has been associated with various adverse events and even IOL surface abnormalities after IOL implantation. In this study, nozzle tip damage of three different injector models with v-notched nozzle tips was systematically evaluated using our self-developed system – the Heidelberg Score for IOL injector damage. Methods: Nozzle tip damage was categorized into 6 grades: no damage (grade 0), slight scratches (1), deep scratches (2), extensions (3), cracks (4), and bursts (5). Each grade was assigned to a score of 0–5. In each IOL injector group, all IOLs were divided into 2 subgroups based on IOL power: +15 to +21D group and +21 to +26D group. The total scores for each group were the sum of scores for all injectors in this group. Further analysis was performed on the nozzle tip configuration and parameters in each injector model. Results: The median (Q1–Q3) for each injector group in group +15 to +21D was 1.5 (1–2) for Avansee, 4 (3–4) for iSert, and 4 (3–4) for multiSert. A statistically significant difference was found between Avansee and iSert (p < 0.001) as well as between Avansee and multiSert (p < 0.01) in terms of median scores. The median (Q1–Q3) for each injector group in group +21 to +26D was 1.5 (1–2) for Avansee, 4 (4–4) for iSert, and 3 (3–3.75) for multiSert. A statistically significant difference was found between Avansee and iSert in terms of median scores (p < 0.001). The outer cross-sectional diameters were 1.80 and 1.78 mm for Avansee, 1.70 and 1.69 mm for iSert, and 1.69 and 1.68 mm for multiSert. The radii of each notch-based circle were 0.21 mm (Avansee), 0.09 (iSert), and 0.06 (multiSert), respectively. The tip angles for three injector models were 48° (Avansee), 46° (iSert), and 37° (multiSert). Conclusions: Avansee showed the least nozzle tip damage of all three groups. Compared with our earlier study using preloaded injectors with intact endpieces, all the injector groups with v-notched nozzle tips had more damage to the nozzle tip. It was found that the closer the notch of the nozzle tip was to the letter “V,” the more damaged the nozzle tip was after IOL implantation.


  • Open Access

    Comparison of a new IOL injector system against 3 standard IOL injector systems with different incision sizes: Miyake-Apple view experimental laboratory study

    by Lu Zhang, Sonja Schickhardt, Hui Fang, Florian Auerbach, Perfecto Iii Cagampang, Patrick R. Merz, Gerd U. Md Febo Auffarth

    Purpose: To compare 1 new intraocular lens injector system against 3 standard injector systems in porcine eyes. Setting: David J Apple Center for Vision Research, Department of Ophthalmology, University Hospital Heidelberg, Heidelberg, Germany. Design: In vitro laboratory study. Methods: In 70 porcine eyes, +20.0 diopter IOLs were implanted with the following systems: multiSert, UltraSert, iTec and RayOne, i.e., S1.8 (incision size: 1.8mm), S2.0 (2.0mm), S2.2P (2.2mm, push-mode), S2.2S (2.2mm, screw-mode), U2.2 (2.2mm), iT2.2 (2.2mm) and R2.0 (2.0mm). Corneal incision sizes were measured before and after implantation with a incision gauge set. Ease of use was evaluated using a Likert scale. IOL delivery time and performance were determined based on Miyake-Apple view videos. Results: The incision enlargements were 0.36 ± 0.08mm (S1.8), 0.15 ± 0.07mm (S2.0), 0.17 ± 0.12mm (S2.2P), 0.28 ± 0.10mm (S2.2S), 0.32 ± 0.09mm (U2.2), 0.30 ± 0.08mm (iT2.2) and 0.35 ± 0.11mm (R2.0). Total scores of ease of use were 23.00 (S1.8), 25.00 (S2.0), 29.00 (S2.2P), 26.00 (S2.2S), 26.00 (U2.2), 25.00 (iT2.2) and 24.00 (R2.0). As for the mean delivery time, iT2.2 took the longest time (13.20 ± 3.29 seconds), while S2.2S took the shortest time (4.50 ± 0.71 seconds). Haptic-optic adhesion was observed in S1.8 (4, 40%), S2.2P (2, 20%), U2.2 (5, 50%) and iT2.2 (5, 50%). Conclusions: Injector S, with the appropriate incision size and implantation method, could achieve better results regarding incision enlargement, ease of use, delivery time and performance than other injector systems. There was an indirect relationship between incision size and inadvertent events.