Tyler Whitaker
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Research Subjects
Tyler Whitaker has a total of 11 co-authors. Here are the top co-authors by number of shared publications:
- Ivan Cardenas
Department of Ophthalmology and Visual Sciences, John A Moran Eye Center, University of Utah, Salt Lake City, UT, USA
2 shared publications - Jeff H. Pettey
Department of Ophthalmology and Visual Sciences, John A Moran Eye Center, University of Utah, Salt Lake City, UT, USA
2 shared publications - Randall J. Olson
Department of Ophthalmology and Visual Sciences, John A Moran Eye Center, University of Utah, Salt Lake City, UT, USA
2 shared publications - Reiker G. Ricks
Department of Ophthalmology and Visual Sciences, John A Moran Eye Center, University of Utah, Salt Lake City, UT, USA
2 shared publications - Tanner K. Nelson
Department of Ophthalmology and Visual Sciences, John A Moran Eye Center, University of Utah, Salt Lake City, UT, USA
2 shared publications Brian C. Stagg
John A. Moran Eye Center, Department of Ophthalmology & Visual Sciences, University of Utah Health
1 shared publication
Latest Publications
- 1 October 2026
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 52 (10) , 1044-1046
Incidence of posterior capsular rupture in routine and complex cataract surgeries at a major training institution
Reports on posterior capsular rupture (PCR) incidence in the US from the past decade are limited, and impact of trainee presence versus responsibility is unclear. We retrospectively analyzed a database of surgical complications within all cataract surgeries from 2022 to 2024 at the Moran Eye Center. PCR rates were stratified by case complexity and presence and direct responsibility of trainees. Among 14,240 cataract surgeries, the PCR rate was 0.67% (95% CI: 0.54-0.81%). PCR occurred more frequently in complex (1.4%) versus routine (0.5%) cases (p<0.001). In cases with PCR, 68% had vitreous loss and 34% reported anterior vitrector use. With assisting trainees present, PCR rate was elevated at 0.94% versus 0.23% without (p 0.40% (p=0.03). Overall, up-to-date PCR incidence from a large academic center is provided. Case complexity and both trainee presence and responsibility significantly increased risk.
- 1 October 2026
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 52 (10) , 1044-1046
Incidence of posterior capsular rupture in routine and complex cataract surgeries at a major training institution
Reports on posterior capsular rupture (PCR) incidence in the US from the past decade are limited, and impact of trainee presence versus responsibility is unclear. We retrospectively analyzed a database of surgical complications within all cataract surgeries from 2022 to 2024 at the Moran Eye Center. PCR rates were stratified by case complexity and presence and direct responsibility of trainees. Among 14,240 cataract surgeries, the PCR rate was 0.67% (95% CI: 0.54-0.81%). PCR occurred more frequently in complex (1.4%) versus routine (0.5%) cases (p<0.001). In cases with PCR, 68% had vitreous loss and 34% reported anterior vitrector use. With assisting trainees present, PCR rate was elevated at 0.94% versus 0.23% without (p 0.40% (p=0.03). Overall, up-to-date PCR incidence from a large academic center is provided. Case complexity and both trainee presence and responsibility significantly increased risk.
- Open Access
- 1 January 2025
- Published by Wiley in Journal of Ophthalmology
- Vol. 2025 (1) , 3264880
Comparison of Postocclusion Pressure Surge Between Pressure Sensing and Traditional Phacoemulsification Handpieces
Purpose: To compare the Active Sentry and OZil handpieces regarding their ability to prevent postocclusion surge. Setting: John A. Moran Eye Center, University of Utah, Salt Lake City, Utah. Design: Experimental study. Methods: This study used the Alcon Centurion surgical platform, with both OZil handpieces fitted with balanced tips. The quad preset was used with a vacuum of 500 mmHg, an aspiration flow of 50 mmHg, and an intraocular pressure (IOP) of 70 mmHg. A rubber disk was fixed within a sealed chamber fitted with an electric pressure sensor to monitor pressure changes. The phaco tip was inserted through a small opening, and the foot pedal was set to position two. The tip was put in contact with the fixed rubber disk to replicate tip occlusion and then pulled from the disk to simulate occlusion break. 10 trials were performed with each handpiece, and pressure changes were recorded electronically. Results: A significant difference ( p < 0.01) was found between the magnitude of postocclusion surge, as well as between the duration of postocclusion surge. Conclusion: The Active Sentry handpiece was found to have a decreased magnitude and shorter duration of postocclusion surge compared with the OZil, signifying that having the pressure sensor in the handpiece allows it to react to post‐occlusion surge more quickly and to decrease surge magnitude.
- 1 January 2025
- Published by Wiley in Journal of Ophthalmology
- Vol. 2025 (1) , 3264880
Open AccessComparison of Postocclusion Pressure Surge Between Pressure Sensing and Traditional Phacoemulsification Handpieces
Purpose: To compare the Active Sentry and OZil handpieces regarding their ability to prevent postocclusion surge. Setting: John A. Moran Eye Center, University of Utah, Salt Lake City, Utah. Design: Experimental study. Methods: This study used the Alcon Centurion surgical platform, with both OZil handpieces fitted with balanced tips. The quad preset was used with a vacuum of 500 mmHg, an aspiration flow of 50 mmHg, and an intraocular pressure (IOP) of 70 mmHg. A rubber disk was fixed within a sealed chamber fitted with an electric pressure sensor to monitor pressure changes. The phaco tip was inserted through a small opening, and the foot pedal was set to position two. The tip was put in contact with the fixed rubber disk to replicate tip occlusion and then pulled from the disk to simulate occlusion break. 10 trials were performed with each handpiece, and pressure changes were recorded electronically. Results: A significant difference ( p < 0.01) was found between the magnitude of postocclusion surge, as well as between the duration of postocclusion surge. Conclusion: The Active Sentry handpiece was found to have a decreased magnitude and shorter duration of postocclusion surge compared with the OZil, signifying that having the pressure sensor in the handpiece allows it to react to post‐occlusion surge more quickly and to decrease surge magnitude.
- Open Access
- 1 January 2025
- Published by Taylor & Francis in Medical Devices: Evidence and Research
- Vol. ume 18, 29-35
Comparison of Ultrasound Energy Delivered to the Anterior Segment Across Different Phacoemulsification Surgical Platforms
Purpose: The aim of this study was to use calorimetry to understand the difference in energy transferred by three phacoemulsification surgical platforms to the eye. Patients and Methods: A phacoemulsification tip was lowered into a double-walled calorimeter filled with distilled water. The foot pedal was depressed for 30 seconds and the change in temperature of the water was measured by a temperature probe. Three phacoemulsification systems were compared: the Alcon Centurion, Johnson & Johnson Veritas and Oertli CataRhex 3. The following conditions remained constant across trials and platforms: continuous longitudinal ultrasound, flow rate 12mL/min, vacuum 0mmHg, and clamped inflow and outflow tubing. The different platforms were directly compared at 20%, 40%, 60%, 80% and 100% power. Results: A two-way ANOVA found a significant difference (P < 0.001) in overall energy output across all trials between the CataRhex 3, Centurion and Veritas with an F value of 63.97 and two degrees of freedom. Conclusion: Given identical settings, the amount of energy produced was significantly different across phacoemulsification platforms. This data can aid surgeons’ understanding of how power level by surgical platform can impact the amount of energy introduced into the anterior segment during cataract surgery. Plain Language Summary: Cataract surgery is one of the most commonly performed surgical procedures in the United States. The most common surgical technique used to remove cataracts is called phacoemulsification (phaco), which utilizes ultrasound energy to break up and remove cataracts. Utilizing phaco to emulsify cataracts is safe and effective; however, because of the delicate nature of the structures in the eye, the amount of ultrasound energy delivered during the surgery is an important consideration to avoid adverse outcomes. There are a variety of phaco platforms (machines) used for cataract surgery. While it is beneficial for surgeons to have many options of surgical platforms, proprietary engineering makes it difficult to know how these surgical platforms compare to one another, especially in important variables such as energy delivery to the eye. To date, it has been unknown how the amount of energy delivered by these phaco platforms differs, if at all, when using identical settings during surgery. We did this study to find out if there are energy differences. To do this research, our lab used calorimetry, which is a way of measuring energy transfer, to compare three different surgical platforms at the same settings. We found a statistically significant difference in amount of energy between these surgical platforms. These results are important because they may help surgeons to understand that the amount of energy delivered during cataract surgery will be different depending on which surgical platform they are using and help to prevent adverse outcomes caused by delivering too much energy to the eye.
- 1 January 2025
- Published by Taylor & Francis in Medical Devices: Evidence and Research
- Vol. ume 18, 29-35
Open AccessComparison of Ultrasound Energy Delivered to the Anterior Segment Across Different Phacoemulsification Surgical Platforms
Purpose: The aim of this study was to use calorimetry to understand the difference in energy transferred by three phacoemulsification surgical platforms to the eye. Patients and Methods: A phacoemulsification tip was lowered into a double-walled calorimeter filled with distilled water. The foot pedal was depressed for 30 seconds and the change in temperature of the water was measured by a temperature probe. Three phacoemulsification systems were compared: the Alcon Centurion, Johnson & Johnson Veritas and Oertli CataRhex 3. The following conditions remained constant across trials and platforms: continuous longitudinal ultrasound, flow rate 12mL/min, vacuum 0mmHg, and clamped inflow and outflow tubing. The different platforms were directly compared at 20%, 40%, 60%, 80% and 100% power. Results: A two-way ANOVA found a significant difference (P < 0.001) in overall energy output across all trials between the CataRhex 3, Centurion and Veritas with an F value of 63.97 and two degrees of freedom. Conclusion: Given identical settings, the amount of energy produced was significantly different across phacoemulsification platforms. This data can aid surgeons’ understanding of how power level by surgical platform can impact the amount of energy introduced into the anterior segment during cataract surgery. Plain Language Summary: Cataract surgery is one of the most commonly performed surgical procedures in the United States. The most common surgical technique used to remove cataracts is called phacoemulsification (phaco), which utilizes ultrasound energy to break up and remove cataracts. Utilizing phaco to emulsify cataracts is safe and effective; however, because of the delicate nature of the structures in the eye, the amount of ultrasound energy delivered during the surgery is an important consideration to avoid adverse outcomes. There are a variety of phaco platforms (machines) used for cataract surgery. While it is beneficial for surgeons to have many options of surgical platforms, proprietary engineering makes it difficult to know how these surgical platforms compare to one another, especially in important variables such as energy delivery to the eye. To date, it has been unknown how the amount of energy delivered by these phaco platforms differs, if at all, when using identical settings during surgery. We did this study to find out if there are energy differences. To do this research, our lab used calorimetry, which is a way of measuring energy transfer, to compare three different surgical platforms at the same settings. We found a statistically significant difference in amount of energy between these surgical platforms. These results are important because they may help surgeons to understand that the amount of energy delivered during cataract surgery will be different depending on which surgical platform they are using and help to prevent adverse outcomes caused by delivering too much energy to the eye.