Assessment of subjective refraction with a clinical adaptive optics visual simulator
Open Access
- 1 January 2019
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 45 (1) , 87-93
Abstract
To clinically validate an adaptive optics visual simulator (VAO) that measures subjective refraction and visual acuity. Optics Laboratory, University of Murcia, Murcia, Spain. Prospective case series. Using the adaptive optics visual simulator, 2 examiners measured the subjective refraction and visual acuity in healthy eyes of volunteers; 1 examiner also used a trial frame as a gold standard. The interexaminer reproducibility and agreement with the gold standard were estimated using the following statistical parameters: limits of agreement from Bland-Altman analysis, significance between differences (P value), and intraclass correlation coefficient (ICC). Seventy-six eyes of 38 volunteers were measured. Interexaminer reproducibility for subjective refraction was excellent (ICC ≥0.96; P > .05), with low 95% confidence interval (CI) values for the power vectors M (spherical equivalent of the given refractive error), J0 (Jackson cross-cylinder, axes at 180 degrees and 90 degrees), and J45 (Jackson cross-cylinder, axes at 45 degrees and 135 degrees) (±0.51 diopter [D], ±0.14 D, and ±0.14 D, respectively). No significant differences in subjective refraction and visual acuity were found between the visual simulator and gold standard (P > .05), with 95% CIs for M, J0, and J45 (subjective refraction) of ±0.67 D, ±0.14 D, and ±0.16 D, respectively, and a ±0.10 logarithm of the minimum angle of resolution (visual acuity). Subjective refraction results using the adaptive optics visual simulator agreed with those of the gold standard and can be used as the baseline for visual simulation of any optical corneal profile or intraocular lens design for refractive surgery patients.Keywords
Funding Information
- European Research Council (ERC-2013-AdG-339228)
- SEIDI (FIS2013-41237-R)
This publication has 33 references indexed in Scilit:
- Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmologyOphthalmic and Physiological Optics, 2011
- Hybrid adaptive-optics visual simulatorOptics Letters, 2010
- Objective and subjective preoperative refraction techniques for wavefront-optimized and wavefront-guided laser in situ keratomileusisJournal of Cataract & Refractive Surgery, 2009
- Reproducibility of sphero‐cylindrical prescriptions*Ophthalmic and Physiological Optics, 2008
- The Precision of Wavefront Refraction Compared to Subjective Refraction and AutorefractionOptometry and Vision Science, 2007
- Repeatability (test‐retest variability) of refractive error measurement in clinical settingsActa Ophthalmologica Scandinavica, 2006
- Adaptive Optics Simulation of Intraocular Lenses with Modified Spherical AberrationInvestigative Ophthalmology & Visual Science, 2004
- Adaptive optics with a programmable phase modulator: applications in the human eyeOptics Express, 2004
- Analysis of the performance of the Hartmann–Shack sensor in the human eyeJournal of the Optical Society of America A, 2000
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986