Demonstration of multi-point stimulation with AC-driven CMOS chips for retinal prosthesis
Authors:
- Yuki NakanishiYNYuki Nakanishi
- Wisaroot SriitsaranusornWSWisaroot Sriitsaranusorn
- Takaya HattoriTHTakaya Hattori
- Kuang-Chih TsoKTKuang-Chih Tso
- Kenzo ShodoKSKenzo Shodo
- Yasuo TerasawaYTYasuo Terasawa
- Yoshinori SunagaYSYoshinori Sunaga
- Hironari TakeharaHTHironari Takehara
- Makito HarutaMHMakito Haruta
- Hiroyuki TashiroHTHiroyuki Tashiro
- Kiyotaka SasagawaKiyotaka Sasagawa
- Jun OhtaJun Ohta
- 14 February 2024
- journal article
- research article
- Published by IOP Publishing in Japanese Journal of Applied Physics
- Vol. 63 (3) , 03SP22
Abstract
We demonstrated a multi-point stimulation system using CMOS chips designed for distributed implantable retinal prostheses. The CMOS chip integrates current sources and electrode selection circuits. This integration enables stimulation current control from multiple electrodes with minimal wires. In this study, we constructed a validation system with 49 electrodes capable of visually observing stimulation current outputs. The validation results show that it is possible to configure stimulation conditions in approximately 9.3 μs per electrode, and we demonstrated the ability to complete one frame, including the stimulation time, within approximately 11 ms. We also demonstrated parallel current stimulation from multiple electrodes using an in vitro experiment.Keywords
This publication has 29 references indexed in Scilit:
- Attaining higher resolution visual prosthetics: a review of the factors and limitationsJournal of Neural Engineering, 2013
- Smart electrode array device with CMOS multi-chip architecture for neural interfaceElectronics Letters, 2012
- Spatial profile of suprachoroidal-transretinal stimulation (STS) for retinal prosthesis evaluated by optical imagingNeuroscience Research, 2011
- Chronic Implantation of Newly Developed Suprachoroidal-Transretinal Stimulation Prosthesis in DogsInvestigative Ophthalmology & Visual Science, 2011
- Blind subjects implanted with the Argus II retinal prosthesis are able to improve performance in a spatial-motor taskBritish Journal of Ophthalmology, 2010
- Neural Stimulation and Recording ElectrodesAnnual Review of Biomedical Engineering, 2008
- Fabrication and Validation of Multichip Neural Stimulator for In vivo Experiments toward Retinal ProsthesisJapanese Journal of Applied Physics, 2007
- Evaluation of phosphenes elicited by extraocular stimulation in normals and by suprachoroidal-transretinal stimulation in patients with retinitis pigmentosaGraefe's Archive for Clinical and Experimental Ophthalmology, 2007
- Laboratory investigation of microelectronics-based stimulators for large-scale suprachoroidal transretinal stimulation (STS)Journal of Neural Engineering, 2007
- CMOS LSI-based multichip flexible neural stimulation device with embedded bulk Pt electrodesElectronics Letters, 2007