Image Noise, CNR, and Detectability of Low-Contrast, Low-Attenuation Liver Lesions in a Phantom: Effects of Radiation Exposure, Phantom Size, Integrated Circuit Detector, and Iterative Reconstruction
- 1 August 2016
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 280 (2) , 475-482
Abstract
To assess image noise, contrast-to-noise ratio (CNR) and detectability of low-contrast, low-attenuation liver lesions in a semianthropomorphic phantom by using either a discrete circuit (DC) detector and filtered back projection (FBP) or an integrated circuit (IC) detector and iterative reconstruction (IR) with changes in radiation exposure and phantom size. An anthropomorphic phantom without or with a 5-cm-thick fat-mimicking ring (widths, 30 and 40 cm) containing liver inserts with four spherical lesions was scanned with five exposure settings on each of two computed tomography scanners, one equipped with a DC detector and the other with an IC detector. Images from the DC and IC detector scanners were reconstructed with FBP and IR, respectively. Image noise and lesion CNR were measured. Four radiologists evaluated lesion presence on a five-point diagnostic confidence scale. Data analyses included receiver operating characteristic (ROC) curve analysis and noninferiority analysis. The combination of IC and IR significantly reduced image noise (P < .001) (with the greatest reduction in the 40-cm phantom and at lower exposures) and improved lesion CNR (P < .001). There was no significant difference in area under the ROC curve between detector-reconstruction combinations at fixed exposure for either phantom. Reader accuracy with IC-IR was noninferior at 50% (100 mAs [effective]) and 25% (300 mAs [effective]) exposure reduction for the 30- and 40-cm phantoms, respectively (adjusted P < .001 and .04 respectively). IC-IR improved readers’ confidence in the presence of a lesion (P = .029) independent of phantom size or exposure level. IC-IR improved objective image quality and lesion detection confidence but did not result in superior diagnostic accuracy when compared with DC-FBP. Moderate exposure reductions maintained comparable diagnostic accuracy for both detector-reconstruction combinations. Lesion detection in the 40-cm phantom was inferior at smaller exposure reduction than in the 30-cm phantom. © RSNA, 2016 Online supplemental material is available for this article.Keywords
This publication has 23 references indexed in Scilit:
- Potential of combining iterative reconstruction with noise efficient detector design: aggressive dose reduction in head CTBritish Journal of Radiology, 2015
- Effect of Reduced Radiation Exposure and Iterative Reconstruction on Detection of Low-Contrast Low-Attenuation Lesions in an Anthropomorphic Liver Phantom: An 18-Reader StudyRadiology, 2014
- Integrated Circuit Detector Technology in Abdominal CT: Added Value in Obese PatientsAmerican Journal of Roentgenology, 2014
- Reducing Image Noise in Computed Tomography (CT) ColonographyJournal of Computer Assisted Tomography, 2014
- Electronic Noise in CT Detectors: Impact on Image Noise and ArtifactsAmerican Journal of Roentgenology, 2013
- Coronary artery stent imaging with CT using an integrated electronics detector and iterative reconstructions: First in vitro experienceJournal of Cardiovascular Computed Tomography, 2013
- Stenosis Quantification in Coronary CT AngiographyInvestigative Radiology, 2013
- Achieving Routine Submillisievert CT Scanning: Report from the Summit on Management of Radiation Dose in CTRadiology, 2012
- Image Noise and Liver Lesion Detection With MDCT: A Phantom StudyAmerican Journal of Roentgenology, 2011
- Visualization and Detection-Localization on Computed Tomographic ImagesInvestigative Radiology, 1991