Effect of Strain Relief Loop Position on the RF-Induced Heating of Active Implantable Medical Devices at 1.5-T MRI
- 29 May 2024
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Electromagnetic Compatibility
- Vol. 66 (4) , 1041-1056
Abstract
This article studies the impact of strain relief loops on RF-induced heating of active implantable medical devices (AIMDs) during magnetic resonance imaging (MRI) scans at 1.5 T. Two commercially available AIMDs, a 40-cm lead length peripheral nerve stimulator (PNS) and a 60-cm lead length spinal cord stimulator (SCS), were utilized. In the in vitro study, lead-tip heating was calculated using the transfer function (TF) method and compared to direct measurement results for strain relief loops at various locations. In the human model analysis, RF-induced heating along six PNS and two SCS trajectories within the Duke, Ella, and FATS models was evaluated for strain relief loops positioned at various locations along the implantation trajectories. From the in vitro study, we can see up to a 48.3% reduction in the overall lead-tip heating. Inappropriate placement may result in doubled temperature rises. Significant temperature rise variation (up to 9°C) is observed with loop position, loop diameter, number of turns, and patient landmark. Differentiation of cumulative summation analysis is necessary to comprehend the strain relief loop's effect on RF-induced heating. The findings suggest that strategically placing the loop can effectively mitigate RF-induced heating for AIMDs exposed to 1.5-T MR environments.Keywords
This publication has 39 references indexed in Scilit:
- RF-induced heating of interventional devices at 23.66 MHzMagnetic Resonance Materials in Physics, Biology and Medicine, 2023
- Upper-Extremity Peripheral Nerve StimulatorsJournal of Hand Surgery Global Online, 2022
- Magnetic resonance conditionality of abandoned leads from active implantable medical devices at 1.5 TMagnetic Resonance in Medicine, 2021
- A technique for the reduction of RF‐induced heating of active implantable medical devices during MRIMagnetic Resonance in Medicine, 2021
- Peripheral Nerve Stimulation: A Review of Techniques and Clinical EfficacyPain and Therapy, 2021
- Effect of Device Configuration and Patient's Body Composition on the RF Heating and Nonsusceptibility Artifact of Deep Brain Stimulation Implants During MRI at 1.5T and 3TJournal of Magnetic Resonance Imaging, 2020
- On the development of equivalent medium for active implantable device radiofrequency safety assessmentMagnetic Resonance in Medicine, 2019
- Safety of active catheters in MRI: Termination impedance versus RF‐induced heatingMagnetic Resonance in Medicine, 2018
- Development of a new generation of high-resolution anatomical models for medical device evaluation: the Virtual Population 3.0Physics in Medicine & Biology, 2014
- Evaluation of the RF heating of a generic deep brain stimulator exposed in 1.5 T magnetic resonance scannersBioelectromagnetics, 2012