Soft Dynamic Fluidic Cushion for Pressure Sore Management in Transtibial Prosthetics: A Proof-of-Concept Study
- 1 March 2026
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 73 (3) , 1221-1235
Abstract
Objective: Due to volume fluctuations, bony prominences contacting prosthetic sockets are susceptible to skin breakdown due to constant pressure and shear stresses from lateral and vertical displacements throughout gait. A low-profile unloading cushion that can be controlled with a millisecond response to dynamically provide localized socket fit at the fibular head may offer stress relief during gait. Methods: The study presents a low-profile (<0.5 mm thick) dynamic fluidic cushion that can provide targeted cushioning during gait to mitigate high contact pressures at bony prominences while minimizing duration of pressure and shear. The proof-of-concept study comprised benchtop testing using a residual limb model. Results: The dynamic cushion effectively unloaded high pressure caused by the body weights only during the stance phase at the fibular head for up to 80 kg by mitigating and redistributing contact pressures away from the fibular head to the immediately surrounding soft tissues. Results demonstrated that the fibular head peak contact pressure could be reduced by as much as 94% (98 to 6 kPa) with a response time of 250 milliseconds in benchtop tests, fast enough to turn on during stance and off during swing. Conclusion: The proposed dynamic fluidic cushion has the potential to offer unloading during gait to prevent skin damage at pressure-sensitive spots, notably benefiting individuals with complex limb geometries. Significance: We introduce a new method with the potential to be integrated into prosthetist workflows to locally adjust socket fit (e.g., at the fibular head) using elastomer sheets and a handheld heat press.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada (2015-05317)
- Idea to Innovation (I2IPJ560618-21)
This publication has 71 references indexed in Scilit:
- Smart Transfemoral Prosthetic Socket with Motorized Cable‐Driven SystemAdvanced Intelligent Systems, 2025
- The relationship between residual limb health, motion within the socket, and prosthetic suspensionPM&R, 2022
- Dressings and topical agents for preventing pressure ulcersCochrane Database of Systematic Reviews, 2018
- Effects of diabetic peripheral neuropathy on gait in vascular trans-tibial amputeesClinical Biomechanics, 2018
- The evidence-base for elevated vacuum in lower limb prosthetics: Literature review and professional feedbackClinical Biomechanics, 2016
- Determinants of Skin Problems of the Stump in Lower-Limb AmputeesArchives of Physical Medicine and Rehabilitation, 2009
- Factors Affecting Perioperative Mortality and Wound-Related Complications Following Major Lower Extremity AmputationsAnnals of Vascular Surgery, 2006
- Regional differences in pain threshold and tolerance of the transtibial residual limb: Including the effects of age and interface materialArchives of Physical Medicine and Rehabilitation, 2005
- Computer-aided design and manufacture of an above-knee amputee socketJournal of Biomedical Engineering, 1991
- A synthesis of the factors that contribute to pressure sore formationMedical Hypotheses, 1983