Removal of ring artifacts in microtomography by characterization of scintillator variations
- 13 September 2017
- journal article
- research article
- Published by Optica Publishing Group in Optics Express
- Vol. 25 (19) , 23191-23198
Abstract
Ring artifacts reduce image quality in tomography, and arise from faulty detector calibration. In microtomography, we have identified that ring artifacts can arise due to high-spatial frequency variations in the scintillator thickness. Such variations are normally removed by a flat-field correction. However, as the spectrum changes, e.g. due to beam hardening, the detector response varies non-uniformly introducing ring artifacts that persist after flat-field correction. In this paper, we present a method to correct for ring artifacts from variations in scintillator thickness by using a simple method to characterize the local scintillator response. The method addresses the actual physical cause of the ring artifacts, in contrary to many other ring artifact removal methods which rely only on image post-processing. By applying the technique to an experimental phantom tomography, we show that ring artifacts are strongly reduced compared to only making a flat-field correction.Keywords
Funding Information
- Knut och Alice Wallenbergs Stiftelse
- Vetenskapsrådet (VR)
This publication has 15 references indexed in Scilit:
- Reduction of ring artifacts in CBCT: Detection and correction of pixel gain variations in flat panel detectorsMedical Physics, 2014
- Micro‐CT artifacts reduction based on detector random shifting and fast data inpaintingMedical Physics, 2013
- An improved method for the removal of ring artifacts in high resolution CT imagingEURASIP Journal on Advances in Signal Processing, 2012
- Removal of ring artifacts in micro-CT imaging using iterative morphological filtersSignal, Image and Video Processing, 2010
- Stripe and ring artifact removal with combined wavelet—Fourier filteringOptics Express, 2009
- Reduction of ring artefacts in high resolution micro-CT reconstructionsPhysics in Medicine & Biology, 2004
- Limitations to flat-field correction methods when using an X-ray spectrumNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Numerical removal of ring artifacts in microtomographyReview of Scientific Instruments, 1998
- X-ray microtomography scanner using time-delay integration for elimination of ring artefacts in the reconstructed imageNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997
- Calculation of Modulation Transfer Functions of X-Ray Fluorescent ScreensApplied Optics, 1973