Abstract
We investigated the impact of measurement circuit configuration, sampler gain error, sampling trigger delay setting, and the built-in filter of the sampler on the accuracy of quantumbased differential sampling. In addition, we evaluated through numerical simulations how the filter function for NI PXI-5922, which changes with sampling rate, is reflected in differential sampling. We found that the filter introduces ripples throughout the differential signal, resulting in a measurement error proportional to frequency. At 3.125 kHz, the maximum difference among the RMS amplitude values obtained with different sampling rates was approximately 2 μV/V. Finally, we performed differential sampling on 1 V AC waveforms using NI PXI-5922 and Fluke 8588A, with bandwidths set to 3 MHz or higher. We found that the RMS amplitude values obtained by the two different samplers agreed within ±52 nV in the frequency range below 1.25 kHz.