Impact of femoral stem alignment on periprosthetic bone density in THA: a study of the Avenir Complete stem
Authors:
- 21 December 2024
- journal article
- research article
- Published by Springer Nature in Archives of orthopaedic and trauma surgery
- Vol. 145 (1) , 1-7
Abstract
Introduction Thigh pain, aseptic loosening, and failure after total hip arthroplasty has been reported. Therefore, this study examines the impact of femoral stem alignment on periprosthetic bone mineral density (BMD) in THA using the Avenir Complete cementless stem, focusing on the role of precise stem alignment in maintaining proximal femoral BMD. Materials and Methods Consecutive patients who received the Avenir Complete stem via mini anterolateral approach in the supine position between March 2019 and March 2022 were included. Dual-energy X-ray absorptiometry was used to assess BMD changes, and computed tomography scans were used to evaluate stem alignment. Results BMD was maintained in the distal femur, but showed a slight decrease in the proximal femur. A negative correlation was observed among varus malalignment, anteversion errors, and changes in BMD in the proximal femur. Conclusions The findings underscore the importance of accurate stem positioning in THA, particularly in avoiding varus malalignment and anteversion errors, as misalignment can adversely affect proximal femoral BMD and potentially lead to early stem loosening. This study enhances the understanding of the influence of the Avenir Complete stem on postoperative BMD, highlighting the need for precise alignment to optimize outcomes.Keywords
This publication has 33 references indexed in Scilit:
- The influence of stem alignment on the bone mineral density around the Polarstem following total hip arthroplastyArchives of orthopaedic and trauma surgery, 2022
- Varus malalignment of short femoral stem not associated with post-hip arthroplasty fractureArchives of orthopaedic and trauma surgery, 2021
- Hydroxyapatite-coated compaction short stem represents a characteristic pattern of peri-prosthetic bone remodelling after total hip arthroplastyArchives of orthopaedic and trauma surgery, 2021
- Migration characteristics of the Corail hydroxyapatite-coated femoral stem—a retrospective clinical evaluation and migration measurement with EBRAArchives of orthopaedic and trauma surgery, 2021
- Risk factors of thigh pain following total hip arthroplasty with short, tapered-wedge stemInternational Orthopaedics, 2020
- Achievement of optimal implant alignment using taper wedge stems with cup-first THA through the MIS antero-lateral approachEuropean Journal of Orthopaedic Surgery & Traumatology, 2020
- Long-term migration characteristics of the Corail hydroxyapatite-coated femoral stem: a 14-year radiostereometric analysis follow-up studyArchives of orthopaedic and trauma surgery, 2019
- Impact of Femoral Stem Design on Failure After Anterior Approach Total Hip ArthroplastyThe Journal of Arthroplasty, 2018
- Periprosthetic bone mineral density with a cementless triple tapered stem is dependent on daily activityInternational Orthopaedics, 2011
- The operation of the century: total hip replacementThe Lancet, 2007