Longevity decoded: Insights from power consumption analyses into device construction and their clinical implications
Author:
Open Access
- 25 February 2019
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 42 (4) , 407-422
Abstract
Introduction: The longevity of a cardiac implantable electronic device (CIED) depends on how quickly the powers consumed by the device's functions exhaust its usable battery energy. A mathematical model for CIED power consumptions was developed and validated against longevity data from manufacturers.Methods: The programmable parameters for the Resonate X4 cardiac resynchronization therapy defibrillators (CRT‐Ds) on the Boston Scientific (St. Paul, MN, USA) online longevity calculator were designated as independent terms in the sum for the total power consumption. The reciprocal of longevity was plotted against variations in these terms. Linear and nonlinear regression analyses were used to fit the plots. The power consumed by pacing was theoretically derived and used as the calibrating tool for estimating the powers consumed by other functions and the usable battery energy. The same methodology was applied to the longevity data of other manufacturers’ CRT‐Ds.Results: Single chamber 100% pacing at 60 beats/min, 2.5 V, 0.4 ms, 500 Ω consumes ≈ 144 J/year. Shock therapy is 45–85% energy efficient. Multichamber pacing modes and maintaining readiness to pace a chamber consume power even if no pacing is delivered. Switching voltage regulation is theoretically more energy efficient than linear voltage regulation for powering pacing.Conclusions: The powers consumed by therapy functions are dictated by the patient's clinical needs, but healthcare professionals can extend device longevity by switching off dormant functions and simplifying the pacing mode. Choosing a device model with large usable battery energy, low background power, and energy efficient pacing and shock therapy for implantation will increase the probability of a long service lifespan.Keywords
This publication has 25 references indexed in Scilit:
- Longevity of implantable cardioverter-defibrillators for cardiac resynchronization therapy in current clinical practice: an analysis according to influencing factors, device generation, and manufacturerEP Europace, 2015
- Novel mechanism of premature battery failure due to lithium cluster formation in implantable cardioverter-defibrillatorsHeart Rhythm, 2014
- Patients treated with catheter ablation for ventricular tachycardia after an ICD shock have lower long-term rates of death and heart failure hospitalization than do patients treated with medical management onlyHeart Rhythm, 2014
- Battery longevity in cardiac resynchronization therapy implantable cardioverter defibrillatorsEP Europace, 2013
- Impact of extending device longevity on the long-term costs of implantable cardioverter-defibrillator therapy: a modelling study with a 15-year time horizonEP Europace, 2013
- Significant differences in the expected versus observed longevity of implantable cardioverter defibrillators (ICDs)Clinical Research in Cardiology, 2012
- Cardiovascular Implantable Electronic Device Replacement Infections and Prevention: Results from the REPLACE RegistryPacing and Clinical Electrophysiology, 2011
- Complication Rates Associated With Pacemaker or Implantable Cardioverter-Defibrillator Generator Replacements and Upgrade ProceduresCirculation, 2010
- Recurrent Implantable Cardioverter-Defibrillator Replacement Is Associated with an Increasing Risk of Pocket-Related ComplicationsPacing and Clinical Electrophysiology, 2010
- Longevity of implantable cardioverter-defibrillators, influencing factors, and comparison to industry-projected longevityHeart Rhythm, 2009