Evaluation of modified window-based scatter compensation in quantitative 177Lu-SPECT for a ring-configured CZT SPECT-CT
Preprint
- 19 January 2026
- preprint
- Published by Springer Nature in Research Square
Abstract
Background : The aim was to evaluate bias and precision for 177 Lu-activity-concentration estimation for window-based scatter compensation in a ring-configured CZT gamma camera. This paper extends a previous study by applying modified scatter compensation (MSC) methods. Whilst the original compensation addressed tailing in the 208 keV peak only, separating primary and scatter photons within the scatter window, the current adaption extents the concept to the 113 keV peak. In the modified version, the estimated primary signal in the scatter window can optionally also be added to measured projections to boost the signal-to-noise ratio (MSC + P). Methods : No additional measurements were acquired, and only previously collected listmode-data for a uniform cylindrical phantom, an image quality NEMA phantom and an anthropomorphic phantom were reframed to accommodate the updated window settings recommended for MSC and MSC + P. Reconstructions were performed using OS-EM with two to 30 iterations (10 subsets) with compensation for attenuation, scatter (DEW/TEW, MSC and MSC + P), distance-dependent spatial resolution, and penetration at 208 keV. Volume-of-interest following the physical sphere size were defined and activity-concentration estimates for each sphere, a liver region and for total activity in the phantoms were assessed in terms of bias and precision for short (10 min) and long timeframes (60 min). Results : Imaging at 208 keV generally result in similar bias and precision and total activity estimates for all scatter-compensation methods. For 113 keV, a slightly netter precision is obtained for MSC and MSC + P but at the cost of larger bias compared with TEW. The major difference is seen for total activity where the MSC and MSC + P manages to decrease bias considerably. The MSC + P yields a decrease in bias of about 10 percentage points when comparing short versus long time frames for activity concentration estimates at 113 keV. Conclusions : Despite the lack of broader improvement in bias and precision when using MSC and MSC + P the findings suggest potential value in more targeted applications. For instance, MSC and MSC + P improves preservation of total activity. Although the findings indicate that using MSC + P is not advisable for estimating high activity concentrations, it may positively influence the estimation of low activity concentrations.Keywords
All Related Versions
- Version 1, 2026-01-19, Research Square
- Published version: EJNMMI Physics, 13 (1).
This publication has 10 references indexed in Scilit:
- Bias and precision of SPECT-based 177Lu activity-concentration estimation using a ring-configured solid-state versus a dual-headed anger systemEJNMMI Physics, 2024
- Quantitative 177Lu SPECT/CT imaging for personalized dosimetry using a ring-shaped CZT-based cameraEJNMMI Physics, 2023
- SPECT at the speed of PET: a feasibility study of CZT-based whole-body SPECT/CT in the post 177Lu-DOTATATE and 177Lu-PSMA617 settingEuropean Journal of Nuclear Medicine and Molecular Imaging, 2023
- Characterisation of a hand-held CZT-based gamma camera for 177Lu imagingEJNMMI Physics, 2020
- Digital Solid-State SPECT/CT Quantitation of Absolute 177Lu Radiotracer Concentration: In Vivo and In Vitro ValidationJournal of Nuclear Medicine, 2020
- Pharmacokinetic digital phantoms for accuracy assessment of image-based dosimetry in177Lu-DOTATATE peptide receptor radionuclide therapyPhysics in Medicine & Biology, 2015
- Scatter correction improves concordance in SPECT MPI with a dedicated cardiac SPECT solid-state cameraJournal of Nuclear Cardiology, 2015
- Initial evaluation of a modified dual-energy window scatter correction method for CZT-based gamma cameras for breast SPECT2015
- MIRD Pamphlet No. 23: Quantitative SPECT for Patient-Specific 3-Dimensional Dosimetry in Internal Radionuclide TherapyJournal of Nuclear Medicine, 2012
- A practical method for position-dependent Compton-scatter correction in single photon emission CTIEEE Transactions on Medical Imaging, 1991