Non-Invasive Cardiac Output Measurement in Low and Very Low Birth Weight: A Method Comparison
Open Access
- 25 March 2014
- journal article
- research article
- Published by Frontiers Media SA in Frontiers in Pediatrics
- Vol. 2, 16
Abstract
Background : Cardiac output (CO) measurement in low (LBW) and very low (VLBW) birth weight infants is difficult. Hitherto, sporadical transthoracic echocardiography (TTE) is the only non-invasive measurement method. Electrical velocimetry (EV) has been evaluated as an alternative in normal weight newborns. Objectives: The study was designed to evaluate if EV could even be interchangeable with TTE in LBW and VLBW infants. Methods: In 28 (17 LBW, 11 VLBW) premature newborns, n = 228 simultaneous TTE (transaortic Doppler) and EV measurements (134 LBW, 94 VLBW) of stroke volume (SV) and heart rate (HR) were performed, thereof calculating body weight indexed SV (= SV*) and CO (= CO*) for all patients and the subgroups. Method comparison was performed by Bland-Altman plot, method precision expressed by calculation of the coefficient of variation (CV). Results: Mean CO* in all patients was 256.4 ± 44.8 (TTE) and 265.3 ± 48.8 (EV) ml/kg/min. Bias and precision were clinically acceptable, limits of agreement within the 30% criterion for method interchangeability (Critchley and Critchley, 1999). According to their different anatomic dimensions and pathophysiology, there were significant differences of SV(*), HR and CO* for LBW and VLBW infants as well for inotropic treatment and ventilation mode.Conclusions: Extending recent publications on EV/TTE comparison in newborns, this study suggests that EV is also applicable in LWB/VLBW infants as a safe and easy to handle method for continuous CO monitoring in the NICU and PCICU.Keywords
This publication has 19 references indexed in Scilit:
- Continuous non-invasive cardiac output measurements in the neonate by electrical velocimetry: a comparison with echocardiographyArchives of Disease in Childhood - Fetal and Neonatal Edition, 2012
- Management of Neonates With Suspected or Proven Early-Onset Bacterial SepsisPediatrics, 2012
- Electrical velocimetry as a tool for measuring cardiac output in small infants after heart surgeryIntensive Care Medicine, 2012
- Near infrared spectroscopy in children at high risk of low perfusionCurrent Opinion in Anaesthesiology, 2010
- Continuous, non-invasive techniques to determine cardiac output in children after cardiac surgery: evaluation of transesophageal Doppler and electric velocimetryJournal of Clinical Monitoring and Computing, 2008
- A Measure of Confidence in Bland–Altman Analysis for the Interchangeability of Two Methods of MeasurementJournal of Clinical Monitoring and Computing, 2008
- Comparison of electrical velocimetry and thermodilution techniques for the measurement of cardiac outputActa Anaesthesiologica Scandinavica, 2007
- Clinicians' abilities to estimate cardiac index in ventilated children and infantsArchives of Disease in Childhood, 1997
- Reproducibility of measurements of cardiac output in newborn infants by Doppler ultrasound.Archives of Disease in Childhood, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986