INVESTIGATION OF THE TL CHARACTERISTICS OF COMPOSITE PEAK 5 IN THE GLOW CURVE OF LIF:MG,TI (TLD-100) USING NATURALLY AND FURNACE-COOLED SAMPLES FOLLOWING THE 400°C PRE-IRRADIATION ANNEAL
- 31 August 2021
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 196 (1-2) , 53-59
Abstract
The cooling rate to room temperature following the 400°C pre-irradiation anneal is known to affect the thermoluminescent properties of LiF:Mg,Ti (TLD-100) as a result of migration and clustering of defects during the cooling down process. In this investigation the dose response over an extended dose range from 0.01 to 7000 Gy in both naturally cooled and the much slower furnace-cooled samples has been measured. Glow curve deconvolution based on first-order kinetics is employed to extract the dose response of the various glow peaks. Of especial interest is the behaviour of glow peaks 4, 5a and 5 as a function of dose. The idea is to modify the supralinear dose response of peak 5 from 1 to 30 Gy to a linear behaviour. This dose range is important for clinical therapy and a linear dose response is of substantial advantage leading to both improved accuracy and precision.Keywords
Funding Information
- PAZY Foundation
- Israeli Ministry of Science and Technology (3-16316, 580298347)
This publication has 20 references indexed in Scilit:
- Thermoluminescence solid-state nanodosimetry--the peak 5A/5 dosemeterRadiation Protection Dosimetry, 2010
- Investigation of the properties of composite glow peak 5 in slow-cooled TLD-100Radiation Measurements, 2008
- The effects of ionisation density on the glow curve structure of LiF:Mg,Ti (TLD-100): the behaviour of composite glow peak 5 in 'slow-cooled' materialRadiation Protection Dosimetry, 2007
- Advanced multistage deconvolution applied to composite glow peak 5 in LiF:Mg,Ti (TLD-100)Radiation Protection Dosimetry, 2007
- Theory of thermoluminescence gamma dose response: The unified interaction modelNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2001
- Investigation of the composite structure of peak 5 in the thermoluminescent glow curve of LiF:Mg,Ti (TLD-100) using optical bleachingJournal of Physics D: Applied Physics, 1999
- A new possible interpretation of the anomalous fading in thermoluminescent materials as normal fading in disguiseRadiation Measurements, 1997
- Editorial - Queer, Quaint, Quirky and Quizzical Thermoluminescent Lithium Fluoride: The 137Co/60Co Calibration Anomaly - A Challenge for NanodosimetryRadiation Protection Dosimetry, 1997
- Effects of cooling and heating rate on trapping parameters in LiF:Mg, Ti crystalsJournal of Physics D: Applied Physics, 1992
- The effect of cooling rate on the performance of thermoluminescent dosimeter crystals (TLD-100) and LiF crystalsJournal of Physics D: Applied Physics, 1976