Calibration of high flow rate thoracic-size selective samplers
Authors:
- 31 March 2016
- journal article
- research article
- Published by Taylor & Francis in Journal of Occupational and Environmental Hygiene
- Vol. 13 (6) , D93-D98
Abstract
High flow rate respirable size selective samplers, GK4.126 and FSP10 cyclones, were calibrated for thoracic-size selective sampling in two different laboratories. The National Institute for Occupational Safety and Health (NIOSH) utilized monodisperse ammonium fluorescein particles and scanning electron microscopy to determine the aerodynamic particle size of the monodisperse aerosol. Fluorescein intensity was measured to determine sampling efficiencies of the cyclones. The Health Safety and Laboratory (HSL) utilized a real time particle sizing instrument (Aerodynamic Particle Sizer) and polydisperse glass sphere particles and particle size distributions between the cyclone and reference sampler were compared. Sampling efficiency of the cyclones were compared to the thoracic convention defined by the American Conference of Governmental Industrial Hygienists (ACGIH)/Comité Européen de Normalisation (CEN)/International Standards Organization (ISO). The GK4.126 cyclone showed minimum bias compared to the thoracic convention at flow rates of 3.5 l min−1 (NIOSH) and 2.7–3.3 l min−1 (HSL) and the difference may be from the use of different test systems. In order to collect the most dust and reduce the limit of detection, HSL suggested using the upper end in range (3.3 l min−1). A flow rate of 3.4 l min−1 would be a reasonable compromise, pending confirmation in other laboratories. The FSP10 cyclone showed minimum bias at the flow rate of 4.0 l min−1 in the NIOSH laboratory test. The high flow rate thoracic-size selective samplers might be used for higher sample mass collection in order to meet analytical limits of quantification.Keywords
This publication has 13 references indexed in Scilit:
- Collection Efficiencies of High Flow Rate Personal Respirable Samplers When Measuring Arizona Road Dust and Analysis of Quartz by X-ray DiffractionAnnals of Work Exposures and Health, 2014
- Evaluation of Pump Pulsation in Respirable Size-Selective Sampling: Part II. Changes in Sampling EfficiencyAnnals of Work Exposures and Health, 2013
- Quartz Measurement in Coal Dust with High-Flow Rate Samplers: Laboratory StudyAnnals of Occupational Hygiene, 2011
- Performance of High Flow Rate Samplers for Respirable Particle CollectionAnnals of Occupational Hygiene, 2010
- Parallel particle impactor – novel size-selective particle sampler for accurate fractioning of inhalable particlesJournal of Physics: Conference Series, 2009
- Measurement of aerosol penetration through six personal thoracic samplers under calm air conditionsJournal of Aerosol Science, 1999
- Development of a system to rapidly measure sampler penetration up to 20 μm aerodynamic diameter in calm air, using the aerodynamic particle sizerJournal of Aerosol Science, 1999
- Characterization and modelling of a family of cyclone aerosol preseparatorsJournal of Aerosol Science, 1997
- A technique for assessing size-selective dust samplers using the APS and polydisperse test aerosolsJournal of Aerosol Science, 1991
- Generation of Large, Solid, Monodisperse Calibration AerosolsAerosol Science and Technology, 1988