Abstract
The potential benefit of organ-based tube current modulation to the female breast in chest CT is overestimated because of a limited reduced tube current zone; despite a 9% dose reduction to the female breast, posterior organs will absorb up to 26% more radiation, resulting in no reduction in radiation-induced malignancies. Purpose To assess the potential dose reduction to the thyroid and breasts in chest computed tomography (CT) with organ-based tube current modulation (OBTCM). Materials and Methods In this retrospective study (from January 2015 to December 2016), the location of the breasts with respect to the reduced tube current zone was determined. With Monte Carlo simulations, patient-specific dose distributions of chest CT scans were calculated for 50 female patients (mean age, 53.7 years ± 17.5; range, 20–80 years). The potential dose reduction with OBTCM was assessed. In addition, simulations of clinical OBTCM scans were made for 17 of the 50 female patients (mean age, 43.8 years ± 17.1; range, 20–69 years). Posterior organs in the field of view were analyzed and lifetime attributable risk (LAR) of cancer incidence and mortality was estimated. Image quality between standard CT and OBTCM scans was compared. Results No women had all breast tissue within the reduced tube current zone. Dose reductions of 18% in the thyr...