Comparison of theTONOVETPlus®, TonoVet®, and Tono‐Pen Vet™ tonometers in normal cats and cats with glaucoma
Open Access
- 20 June 2023
- journal article
- research article
- Published by Wiley in Veterinary Ophthalmology
- Vol. 26 (5) , 414-421
Abstract
Objectives: To determine the accuracy, precision, and clinical applicability of the ICare® TONOVET Plus (TVP) in cats.Animals and Procedures: IOP readings obtained with the TVP were compared to values obtained concurrently with the original TONOVET (TV01) and Tono‐Pen Vet™ (TP) in 12 normal cats (24 eyes) and 8 glaucomatousLTBP2‐mutant cats (13 eyes) in vivo. Reproducibility of TVP readings was also assessed for three observers in the above cats. The anterior chambers of five different normal cat eyes were cannulated ex vivo. IOP was measured with the TVP, TV01, and TP at manometric IOPs ranging from 5 to 70 mmHg. Data were analyzed by linear regression, ANOVA and Bland–Altman plots. ANOVA was used to assess reproducibility of TVP readings obtained by different observers and an ANCOVA model controlled for variation of individual cats.p < .05 was considered significant.Results: TVP values strongly correlated with TV01 values (y = 1.045x + 1.443,R2 = .9667). The TP significantly underestimated IOP relative to the TVP and TV01, particularly at high IOP. IOP values obtained by 1 observer were significantly higher (~1 mmHg average) compared to the other 2 observers via ANCOVA analysis (p = .0006479 andp = .0203). Relative to manometry, the TVP and TV01 were significantly more accurate (p < .0001) and precise (p < .0070) than the TP in ex vivo eyes.Conclusions: IOP readings obtained with the TVP and TV01 are broadly interchangeable between models and between observers, but subtle differences may be important in a research context. TP readings vastly underestimate high IOP in feline glaucoma.Keywords
Funding Information
- National Eye Institute (P30 EY16665)
- Research to Prevent Blindness
This publication has 36 references indexed in Scilit:
- Validation of the TonoVet®rebound tonometer in normal and glaucomatous catsVeterinary Ophthalmology, 2012
- Accuracy of intraocular pressure measurements in dogs using two different tonometers and plano therapeutic soft contact lensesVeterinary Ophthalmology, 2011
- Feline glaucoma—a comprehensive reviewVeterinary Ophthalmology, 2011
- Evaluation of rebound tonometry in non-human primatesExperimental Eye Research, 2011
- Accuracy and reproducibility of the TonoVet®rebound tonometer in birds of preyVeterinary Ophthalmology, 2010
- Intraocular pressure measurement in mice: a comparison between Goldmann and rebound tonometryEye, 2006
- Comparison of the use of new handheld tonometers and established applanation tonometers in dogsAmerican Journal of Veterinary Research, 2006
- Ex vivo Evaluation of Tono-Pen and Pneumotonometry in Cat EyesOphthalmic Research, 2005
- The feline glaucomas: 82 cases (1995–1999)Veterinary Ophthalmology, 2001
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986