Abstract
The shrinkage and deformation of the mesh can be the causes of hernia recurrence and other related complications. Accurate visualization of the mesh and its position would greatly aid in assessing mesh shrinkage, evaluating the risk of hernia recurrence, and setting the preoperative plan for repair. The purpose of this study is to noninvasively describe and quantify the shrinkage, movement, and centroid position of lightweight mesh implants using Automated 3D breast ultrasound (ABUS) through in vivo and vitro studies. In vitro study, ABUS can visualize spatial morphological changes in meshes, which solves the problem of difficult imaging detection of lightweight mesh. The vitro study showed that the maximum deviation between the quantification based on ABUS and the actual mesh size was 3.5%. In vivo study, ABUS not only allows for imaging conventional lightweight meshes but also for assessing time-dependent changes in lightweight mesh implants. The average shrinkage from day 5 to day 90 after surgery was 20.2%, while the average deviation in centroid position was 0.23cm in the vivo study. The problem of postoperative evaluation of lightweight mesh imaging can be effectively addressed by time-dependent assessing and analysis of mesh implants in patients.