ACPSEM position paper: dosimetry for magnetic resonance imaging linear accelerators
Authors:
- Jarrad BeggJBJarrad Begg
- Urszula JelenUJUrszula Jelen
- Zoe MoutrieZMZoe Moutrie
- Chris OliverCOChris Oliver
- Lois HollowayLHLois Holloway
- Rhonda BrownRBRhonda Brown
- the ACPSEM Magnetic Resonance Imaging Linac Working Group (MRILWG)T(the ACPSEM Magnetic Resonance Imaging Linac Working Group (MRILWG)
Open Access
- 20 February 2023
- journal article
- research article
- Published by Springer Nature in Physical and Engineering Sciences in Medicine
- Vol. 46 (1) , 1-17
Abstract
Consistency and clear guidelines on dosimetry are essential for accurate and precise dosimetry, to ensure the best patient outcomes and to allow direct dose comparison across different centres. Magnetic Resonance Imaging Linac (MRI-linac) systems have recently been introduced to Australasian clinics. This report provides recommendations on reference dosimetry measurements for MRI-linacs on behalf of the Australiasian College of Physical Scientists and Engineers in Medicine (ACPSEM) MRI-linac working group. There are two configurations considered for MRI-linacs, perpendicular and parallel, referring to the relative direction of the magnetic field and radiation beam, with different impacts on dose deposition in a medium. These recommendations focus on ion chambers which are most commonly used in the clinic for reference dosimetry. Water phantoms must be MR safe or conditional and practical limitations on phantom set-up must be considered. Solid phantoms are not advised for reference dosimetry. For reference dosimetry, IAEA TRS-398 recommendations cannot be followed completely due to physical differences between conventional linac and MRI-linac systems. Manufacturers’ advice on reference conditions should be followed. Beam quality specification of TPR20,10 is recommended. The configuration of the central axis of the ion chamber relative to the magnetic field and radiation beam impacts the chamber response and must be considered carefully. Recommended corrections to delivered dose are $${k}_{{Q}_{msr}{Q}_{0}}^{{f}_{msr}{f}_{ref}}$$ , a correction for beam quality and $${k}_{\overrightarrow{B},{Q}_{msr}}^{{f}_{msr}}$$ , for the impact of the magnetic field on dosimeter response in the magnetic field. Literature based values for $${k}_{\overrightarrow{B},{Q}_{msr}}^{{f}_{msr}}$$ are given. It is important to note that this is a developing field and these recommendations should be used together with a review of current literature.
Keywords
This publication has 68 references indexed in Scilit:
- Towards reference dosimetry for the MR-linac: magnetic field correction of the ionization chamber readingPhysics in Medicine & Biology, 2013
- ACR guidance document on MR safe practices: 2013Journal of Magnetic Resonance Imaging, 2013
- Electron contamination modeling and skin dose in 6 MV longitudinal field MRIgRT: Impact of the MRI and MRI fringe fieldMedical Physics, 2012
- A study of the effect of in-line and perpendicular magnetic fields on beam characteristics of electron guns in medical linear acceleratorsMedical Physics, 2011
- Effect of transverse magnetic fields on a simulated in-line 6 MV linacPhysics in Medicine & Biology, 2010
- Dosimetry for the MRI accelerator: the impact of a magnetic field on the response of a Farmer NE2571 ionization chamberPhysics in Medicine & Biology, 2009
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: impact of the surface orientation on the entrance and exit dose due to the transverse magnetic fieldPhysics in Medicine & Biology, 2007
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue–air interfaces in a lateral magnetic field due to returning electronsPhysics in Medicine & Biology, 2005
- 1. IntroductionActa Oncologica, 1988
- What degree of accuracy is required and can be achieved in photon and neutron therapy?Radiotherapy and Oncology, 1987