Synthesized V7 QRS Amplitude and Oversensing Episodes in Patients With Subcutaneous Implantable Cardioverter‐Defibrillators
- 5 October 2024
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 47 (12) , 1565-1573
Abstract
Background: Patients with subcutaneous implantable cardioverter‐defibrillators (S‐ICDs) experience an oversensing episode (OS) more frequently than those with transvenous ICDs. However, no established electrocardiography (ECG) parameters can accurately detect an OS. This study aimed to evaluate the incidence of an OS in real‐world clinical practice and the association of synthesized 18‐lead ECG (syn18‐ECG) parameters with an OS.Methods: We retrospectively included 21 consecutive patients who underwent S‐ICD implantation and collected syn18‐ECG parameters. We placed the generator in a deep posterior position and defined an OS as an inappropriate charging episode caused by cardiac or noncardiac signals. A SMART pass filter and two tachyarrhythmia zones were programed.Results: The most frequent underlying heart disease was Brugada/J wave syndrome (n = 7). During a median follow‐up period of 1188 days, an OS was observed in six patients (28.6%). The QRS amplitude in synthesized V7 lead (synV7) was significantly lower in the OS group than in the non‐OS group (0.59 ± 0.17 vs. 0.91 ± 0.35 mV, p = 0.019). The optimal cutoff value of synV7 QRS amplitude was 0.61 mV, with a sensitivity of 80.0% and a specificity of 83.7% for predicting an OS. Univariate logistic analysis showed that a synV7 QRS amplitude of p = 0.018).Conclusions: In patients with S‐ICDs, an OS was not a rare complication during long‐term follow‐up. A low synV7 QRS amplitude was associated with a high OS incidence.Keywords
This publication has 18 references indexed in Scilit:
- Exploiting SMART pass filter deactivation detection to minimize inappropriate subcutaneous implantable cardioverter defibrillator therapies: a real-world single-centre experience and management guideEP Europace, 2023
- Clinical impact of ECG changes on oversensing of subcutaneous implantable cardioverter-defibrillatorsHeart Rhythm, 2022
- JCS/JHRS 2019 guideline on non‐pharmacotherapy of cardiac arrhythmiasJournal of Arrhythmia, 2021
- Oversensing issues leading to device extraction: When subcutaneous implantable cardioverter-defibrillator reached a dead-endHeart Rhythm, 2020
- How the S-ICD (subcutaneous implantable cardiac defibrillator) senses cardiac signals to minimize cardiac over-sensing and maximize rhythm discriminationJournal of Electrocardiology, 2018
- Clinical and electrocardiographic predictors of T wave oversensing in patients with subcutaneous ICDJournal of Arrhythmia, 2016
- Safety and Efficacy of the Totally Subcutaneous Implantable DefibrillatorJACC, 2015
- The QRS morphology pattern in V5R is a novel and simple parameter for differentiating the origin of idiopathic outflow tract ventricular arrhythmiasEP Europace, 2015
- Estimation of the origin of ventricular outflow tract arrhythmia using synthesized right-sided chest leadsEP Europace, 2013
- Prognostic Importance of Defibrillator Shocks in Patients with Heart FailureNew England Journal of Medicine, 2008