MRI‐based transfer function determination through the transfer matrix by jointly fitting the incident and scattered field
Authors:
Open Access
- 21 October 2019
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 83 (3) , 1081-1095
Abstract
Purpose A purely experimental method for MRI‐based transfer function (TF) determination is presented. A TF characterizes the potential for radiofrequency heating of a linear implant by relating the incident tangential electric field to a scattered electric field at its tip. We utilize the previously introduced transfer matrix (TM) to determine transfer functions solely from the MR measurable quantities, that is, the and transceive phase distributions. This technique can extend the current practice of phantom‐based TF assessment with dedicated experimental setup toward MR‐based methods that have the potential to assess the TF in more realistic situations. Theory and Methods An analytical description of the magnitude and transceive phase distribution around a wire‐like implant was derived based on the TM. In this model, the background field is described using a superposition of spherical and cylindrical harmonics while the transfer matrix is parameterized using a previously introduced attenuated wave model. This analytical description can be used to estimate the transfer matrix and transfer function based on the measured distribution. Results The TF was successfully determined for 2 mock‐up implants: a 20‐cm bare copper wire and a 20‐cm insulated copper wire with 10 mm of insulation stripped at both endings in respectively 4 and 3 different trajectories. The measured TFs show a strong correlation with a reference determined from simulations and between the separate experiments with correlation coefficients above 0.96 between all TFs. Compared to the simulated TF, the maximum deviation in the estimated tip field is 9.4% and 12.2% for the bare and insulated wire, respectively. Conclusions A method has been developed to measure the TF of medical implants using MRI experiments. Jointly fitting the incident and scattered distributions with an analytical description based on the transfer matrix enables accurate determination of the TF of 2 test implants. The presented method no longer needs input from simulated data and can therefore, in principle, be used to measure TF's in test animals or corpses.Keywords
Funding Information
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek (15739)
This publication has 25 references indexed in Scilit:
- MRI‐based, wireless determination of the transfer function of a linear implant: Introduction of the transfer matrixMagnetic Resonance in Medicine, 2018
- Toward in vivo quantification of induced RF currents on long thin conductorsMagnetic Resonance in Medicine, 2018
- Experimental setup for transfer function measurement to assess RF heating of medical leads in MRI: Validation in the case of a single wireMagnetic Resonance in Medicine, 2017
- MRI‐based transfer function determination for the assessment of implant safetyMagnetic Resonance in Medicine, 2017
- A Technique to Evaluate MRI-Induced Electric Fields at the Ends of Practical Implanted LeadIEEE Transactions on Microwave Theory and Techniques, 2014
- Safely assessing radiofrequency heating potential of conductive devices using image‐based current measurementsMagnetic Resonance in Medicine, 2014
- Iterative separation of transmit and receive phase contributions and B 1 + -based estimation of the specific absorption rate for transmit arraysMagnetic Resonance Materials in Physics, Biology and Medicine, 2013
- Robust reconstruction of B1+ maps by projection into a spherical functions spaceMagnetic Resonance in Medicine, 2013
- New method to monitor RF safety in MRI‐guided interventions based on RF induced image artefactsMedical Physics, 2010
- Calculation of MRI‐induced heating of an implanted medical lead wire with an electric field transfer functionJournal of Magnetic Resonance Imaging, 2007