Development and evaluation of polycrystalline semiconductor dosimeter for quality assurance application in high-dose-rate brachytherapy
Authors:
- 1 December 2025
- journal article
- Published by IOP Publishing in Journal of Instrumentation
- Vol. 20 (12) , P12005
Abstract
This study aims to develop and evaluate polycrystalline semiconductor dosimeters (PSDs) to establish a digital quality assurance (QA) system for high-dose-rate (HDR) brachytherapy. PSDs were fabricated using four polycrystalline materials: mercury(II) iodide (HgI 2 ), lead(II) iodide (PbI 2 ), lead(II) oxide (PbO), and thallium(I) bromide (TlBr). To evaluate the performance of the fabricated dosimeters, dosimetric parameters, including reproducibility, dose-response linearity, distance dependence, and angular dependence, were measured using an Iridium-192 (Ir-192) radioisotope source and compared with corresponding parameters measured by a parallel-plane ionization chamber. For reproducibility, each PSD was irradiated 10 consecutive times. Dose-response linearity was assessed for doses ranging from 0.1 to 10 Gy. Distance dependence was evaluated by administering a dose of 1 Gy at source-to-dosimeter distances ranging from 1 to 12 cm, in increments of 0.25 cm. Angular dependence was assessed by positioning the PSDs 1 cm below the Ir-192 source and measuring signals at angles of 15°, 30°, and 45° relative to the reference angle of 0°, where the PSD's central axis was parallel to the source. Signals measured at 15°, 30°, and 45° were normalized to those measured at 0°. The PSD using PbO demonstrated the best reproducibility, with a coefficient of variation of 0.85%, and showed a minimal difference of -0.03% compared with the ionization chamber. Other PSDs also achieved excellent reproducibility, with relative standard deviations within 1.5%. For dose-response linearity, all PSDs achieved an R 2 value of 0.9993 or higher, outperforming the ionization chamber, which had an R 2 value of 0.9983. Regarding distance dependence, the slope of the power function and the distance corresponding to 50% intensity ( D 50 ) were analyzed in comparison with the ionization chamber. The PSD using PbI 2 exhibited the closest agreement, with a slope difference of 0.103 and a difference of 0.003 cm for D 50 . For angular dependence, the PSD using PbI 2 demonstrated the least attenuation, with a redcution of 13.2% in signal intensity at 45° compared with 0°. The findings of this study indicated that PbI 2 was the most suitable material for PSDs by comparing with the ionization chamber in terms of measurement characteristics. Furthermore, this study provided valuable insights into optimizing semiconductor materials for various applications in HDR brachytherapy QA.Keywords
This publication has 12 references indexed in Scilit:
- Suitability Assessment of CsPbI_3 Dosimeter with ^192Ir Source for Brachytherapy Quality AssuranceJournal of Instrumentation, 2023
- Development and evaluation of a thallium (I) bromide dosimeter for intracavitary radiotherapy quality assuranceJournal of Instrumentation, 2022
- Automated multi-parameter high-dose-rate brachytherapy quality assurance via radioluminescence imagingPhysics in Medicine & Biology, 2020
- Repeatability and reproducibility of measurements of low dissolved radiocesium concentrations in freshwater using different pre-concentration methodsJournal of Radioanalytical and Nuclear Chemistry, 2019
- Prescribing, Recording, and Reporting Brachytherapy for Cancer of the CervixJournal of the ICRU, 2013
- A Study on the Dose Distributions in Various Materials from an Ir-192 HDR Brachytherapy SourcePLOS ONE, 2012
- A quality assurance tool for high‐dose‐rate brachytherapyMedical Physics, 2010
- High dose rate (HDR) brachytherapy quality assurance: a practical guideBiomedical Imaging and Intervention Journal, 2006
- Reproducibility study of TLD-100 micro-cubes at radiotherapy dose levelApplied Radiation and Isotopes, 1999
- Determination of Absorbed Dose in a Patient Irradiated by Beams of X or Gamma Rays in Radiotherapy ProceduresMedical Physics, 1977