Non-invasive estimation of pressure drop across aortic coarctations: validation of 0D and 3D computational models with in vivo measurements
Preprint
Abstract
Purpose: Blood pressure gradient ( Δ P) across an aortic coarctation (CoA) is an important measurement to diagnose CoA severity and gauge treatment efficacy. Invasive cardiac catheterization is currently the gold-standard method for measuring blood pressure. The objective of this study was to evaluate the accuracy of Δ P estimates derived non-invasively using patient-specific 0D and 3D deformable wall simulations. Methods: Medical imaging and routine clinical measurements were used to create patient-specific models of patients with CoA (N=17). 0D simulations were performed first and used to tune boundary conditions and initialize 3D simulations. Δ P across the CoA estimated using both 0D and 3D simulations were compared to invasive catheter-based pressure measurements for validation. Results: The 0D simulations were extremely efficient ( ∼ 15 secs computation time) compared to 3D simulations ( ∼ 30 hrs computation time on a cluster). However, the 0D Δ P estimates, unsurprisingly, had larger mean errors when compared to catheterization than 3D estimates (12.1 ± 9.9 mmHg vs 5.3 ± 5.4 mmHg). In particular, the 0D model performance degraded in cases where the CoA was adjacent to a bifurcation. The 0D model classified patients with severe CoA requiring intervention (defined as Δ P ≥ 20 mmHg) with 76% accuracy and 3D simulations improved this to 88%. Conclusion: Overall, a combined approach, using 0D models to efficiently tune and launch 3D models, offers the best combination of speed and accuracy for non-invasive classification of CoA severity.Keywords
This publication has 31 references indexed in Scilit:
- Quantification of Local Hemodynamic Alterations Caused by Virtual Implantation of Three Commercially Available Stents for the Treatment of Aortic CoarctationPediatric Cardiology, 2013
- Non-Invasive Hemodynamic Assessment of Aortic Coarctation: Validation with In Vivo MeasurementsAnnals of Biomedical Engineering, 2012
- Computational Simulations for Aortic Coarctation: Representative Results From a Sampling of PatientsJournal of Biomechanical Engineering, 2011
- Evaluation of a novel Y-shaped extracardiac Fontan baffle using computational fluid dynamicsThe Journal of Thoracic and Cardiovascular Surgery, 2009
- ACC/AHA 2008 Guidelines for the Management of Adults With Congenital Heart DiseaseCirculation, 2008
- A coupled momentum method for modeling blood flow in three-dimensional deformable arteriesComputer Methods in Applied Mechanics and Engineering, 2006
- In vivo validation of a one-dimensional finite-element method for predicting blood flow in cardiovascular bypass graftsIEEE Transactions on Biomedical Engineering, 2003
- On the coupling of 3D and 1D Navier–Stokes equations for flow problems in compliant vesselsComputer Methods in Applied Mechanics and Engineering, 2001
- Computer simulation of local blood flow and vessel mechanics in a compliant carotid artery bifurcation modelJournal of Biomechanics, 1995
- Influence of Aortic Coarctation on Pulsatile Hemodynamics in the Proximal AortaCirculation, 1971