Analysis of the deterioration of photostimulable phosphor plates
- 1 September 2020
- journal article
- research article
- Published by Oxford University Press (OUP) in Dentomaxillofacial Radiology
- Vol. 49 (6) , 20190500
Abstract
To assess, objectively to determine the deterioration of the photostimulable phosphor plates (PSP) after several acquisitions. Two new PSPs without preliminary use were exposed with an intra oral X-ray unit, and the EXPRESS® (Instrumentarium Imaging, Tuusula, Finland) unit was used for scanning of the PSPs. A 12-stepwedge aluminum scale were used superimposed with the PSPs during the X-rays exposures to perform the objective analyses. Objective analysis was carried out by mean greys value using the Image J software through a region of interest (ROI) of 0.1 × 0.1 millimeters, and the data were analyzed statistically using ANOVA. For each the Express® PSP a total of 1800 images were acquired during 60 days. The objective analysis showed loss of the mean greys values between the initial and final images. Besides, the percentage of MGVs’ loss was between 0.7%, for the thickest step, and 8.4%, for the lowest thickness step. After many X-ray exposures of the Express® PSP a singular deterioration in the mean greys values could be observed.Keywords
This publication has 10 references indexed in Scilit:
- Assessment of intraoral image artifacts related to photostimulable phosphor plates in a dentomaxillofacial radiology departmentNigerian Journal of Clinical Practice, 2016
- Evaluation of image quality parameters of representative intraoral digital radiographic systemsOral Surgery, Oral Medicine, Oral Pathology and Oral Radiology, 2013
- Durability of Digora Optime® imaging platesOral Radiology, 2011
- Work flow with digital intraoral radiography: A systematic reviewActa Odontologica Scandinavica, 2010
- How many times can we use a phosphor plate? A preliminary studyDentomaxillofacial Radiology, 2009
- How does signal fade on photo-stimulable storage phosphor imaging plates when scanned with a delay and what is the effect on image quality?Oral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2006
- What happens if you delay scanning Digora phosphor storage plates (PSPs) for up to 4 hours?Dentomaxillofacial Radiology, 2006
- Filmless imagingThe Journal of the American Dental Association, 2005
- Effect of delayed scanning of storage phosphor platesOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2005
- Storage Phosphor Plates: How Durable are they as a Digital Dental Radiographic System?The Journal of Contemporary Dental Practice, 2004