Tesla-scale magnetic field measurement based on Sideband-overlap Zeeman spectroscopy using a functionalized MEMS vapor cell
Open Access
- 5 June 2026
- journal article
- Published by Springer Nature in Microsystems & Nanoengineering
- Vol. 12 (1)
Abstract
Accurate measurement of strong magnetic fields in the Tesla range remains a persistent challenge due to calibration drift, nonlinearity, and spatial gradient sensitivity. Here, we present a compact magnetometry approach based on sideband-overlap Zeeman spectroscopy, enabled by a functionalized MEMS cesium vapor cell. The vapor cell features a dual-chamber glass–Si–glass structure with integrated microheaters on the optical window. A first-order electro-optic modulator produces optical sidebands aligned with the σ+ and \documentclass[12pt]{minimal}
\usepackage{amsmath}
\usepackage{wasysym}
\usepackage{amsfonts}
\usepackage{amssymb}
\usepackage{amsbsy}
\usepackage{mathrsfs}
\usepackage{upgreek}
\setlength{\oddsidemargin}{-69pt}
$${{\rm{\sigma }}}^{-}$$ Zeeman-split components of the Cs D1 line under the hyperfine Paschen–Back regime. Magnetic field scanning from zero to Tesla-scale fields enables extraction of the frequency offset between reference and Zeeman-shifted spectra. This frequency shift is directly converted into magnetic field strength using Zeeman spectroscopy in the hyperfine Paschen–Back regime. The method resolves sixteen Zeeman transitions (eight \documentclass[12pt]{minimal}
\usepackage{amsmath}
\usepackage{wasysym}
\usepackage{amsfonts}
\usepackage{amssymb}
\usepackage{amsbsy}
\usepackage{mathrsfs}
\usepackage{upgreek}
\setlength{\oddsidemargin}{-69pt}
$${{\rm{\sigma }}}^{+}$$ and eight \documentclass[12pt]{minimal}
\usepackage{amsmath}
\usepackage{wasysym}
\usepackage{amsfonts}
\usepackage{amssymb}
\usepackage{amsbsy}
\usepackage{mathrsfs}
\usepackage{upgreek}
\setlength{\oddsidemargin}{-69pt}
$${{\rm{\sigma }}}^{-}$$ ), allowing measurement of a field strength of 0.6694921 T, with a single-shot resolution of 6.4 μT and a cross-transition repeatability of σB = 8.9 μT (corresponding to 13 ppm at 0.6695 T). In a millimeter-scale sensing volume, the system achieves sub-10 μT repeatability without requiring a uniform bias field. These results indicate that the microfabricated high-field magnetometer can serve as a compact and practical approach for Tesla-range field measurement and pave the way for chip-scale quantum devices.
Keywords
This publication has 39 references indexed in Scilit:
- The fabrication of MEMS alkali metal vapor cells based on ultrafast laser welding for single beam magnetometerMicrosystems & Nanoengineering, 2025
- Ultrasensitive SERF atomic magnetometer with a miniaturized hybrid vapor cellMicrosystems & Nanoengineering, 2024
- Observation of magnetically-induced transition intensity redistribution in the onset of the hyperfine Paschen–Back regimeOptik, 2024
- Magnetic Resonance Based Atomic Magnetometers2016
- Spectroscopy of Natural and Artificial Atoms in Magnetic Fields2016
- Atomic magnetometers for materials characterizationMaterials Today, 2011
- Optical magnetometryNature Physics, 2007
- Advances in fluxgate sensorsSensors and Actuators A: Physical, 2003
- A subfemtotesla multichannel atomic magnetometerNature, 2003
- Cesium saturation spectroscopy revisited: How to reverse peaks and observe narrow resonancesApplied Physics B, 1994