A temperature sensor implant for active implantable medical devices for in vivo subacute heating tests under MRI
- 14 September 2017
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 79 (5) , 2824-2832
Abstract
Purpose: To introduce a temperature sensor implant (TSI) that mimics an active implantable medical device (AIMD) for animal testing of MRI heating. Computer simulations and phantom experiments poorly represent potential temperature increases. Animal experiments could be a better model, but heating experiments conducted immediately after the surgery suffer from alterations of the thermoregulatory and tissue properties during acute testing conditions. Therefore, the aim of this study was to introduce a temperature sensor implant that mimics an AIMD and capable of measuring the electrode temperature after implantation of the device without any further intervention at any time after the surgery in an animal model.Methods: A battery‐operated TSI, which resembled an AIMD, was used to measure the lead temperature and impedance and the case temperature. The measured values were transmitted to an external computer via a low‐power Bluetooth communication protocol. In addition to validation experiments on the phantom, a sheep experiment was conducted to test the feasibility of the system in subacute conditions.Results: The measurements had a maximum of 0.5°C difference compared to fiber‐optic temperature probes. In vivo animal experiments demonstrated feasibility of the system.Conclusion: An active implant, which can measure its own temperature, was proposed to investigate implant heating during MRI examinations. Magn Reson Med 79:2824–2832, 2018. © 2017 International Society for Magnetic Resonance in Medicine.Keywords
Funding Information
- Turkish Ministry of Science, Industry, and Technology (0204.STZ.2013-1)
This publication has 26 references indexed in Scilit:
- Construction and modeling of a reconfigurable MRI coil for lowering SAR in patients with deep brain stimulation implantsNeuroImage, 2017
- Heating and Safety Concerns of the Radio-Frequency Field in MRICurrent Radiology Reports, 2015
- Parallel transmit excitation at 1.5 T based on the minimization of a driving function for device heatingMedical Physics, 2015
- Controlling radiofrequency-induced currents in guidewires using parallel transmitMagnetic Resonance in Medicine, 2014
- Parallel transmit pulse design for patients with deep brain stimulation implantsMagnetic Resonance in Medicine, 2014
- Safely assessing radiofrequency heating potential of conductive devices using image‐based current measurementsMagnetic Resonance in Medicine, 2014
- Reduction of the radiofrequency heating of metallic devices using a dual‐drive birdcage coilMagnetic Resonance in Medicine, 2012
- Reduction of implant RF heating through modification of transmit coil electric fieldMagnetic Resonance in Medicine, 2010
- Complexity of MRI induced heating on metallic leads: Experimental measurements of 374 configurationsBioMedical Engineering OnLine, 2008
- MRI and implanted medical devices: basic interactions with an emphasis on heatingIEEE Transactions on Device and Materials Reliability, 2005