Dosimetric characterisation of the optically-stimulated luminescence dosimeter in cobalt-60 high dose rate brachytherapy system
- 13 May 2018
- journal article
- research article
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 41 (2) , 475-485
Abstract
This study investigates the characteristics and application of the optically-stimulated luminescence dosimeter (OSLD) in cobalt-60 high dose rate (HDR) brachytherapy, and compares the results with the dosage produced by the treatment planning system (TPS). The OSLD characteristics comprised linearity, reproducibility, angular dependence, depth dependence, signal depletion, bleaching rate and cumulative dose measurement. A phantom verification exercise was also conducted using the Farmer ionisation chamber and in vivo diodes. The OSLD signal indicated a supralinear response (R2 = 0.9998). It exhibited a depth-independent trend after a steep dose gradient region. The signal depletion per readout was negligible (0.02%), with expected deviation for angular dependence due to off-axis sensitive volume, ranging from 1 to 16%. The residual signal of the OSLDs after 1 day bleached was within 1.5%. The accumulated and bleached OSLD signals had a standard deviation of ± 0.78 and ± 0.18 Gy, respectively. The TPS was found to underestimate the measured doses with deviations of 5% in OSLD, 17% in the Farmer ionisation chamber, and 7 and 8% for bladder and rectal diode probes. Discrepancies can be due to the positional uncertainty in the high-dose gradient. This demonstrates a slight displacement of the organ at risk near the steep dose gradient region will result in a large dose uncertainty. This justifies the importance of in vivo measurements in cobalt-60 HDR brachytherapy.Keywords
Funding Information
- University of Malaya Postgraduate Research Fund PPP (PO026-2015B)
This publication has 28 references indexed in Scilit:
- Optically Stimulated Luminescent Dosimetry for High Dose Rate BrachytherapyFrontiers in Oncology, 2012
- Optically stimulated luminescencein vivodosimetry for radiotherapy: physical characterization and clinical measurements in60Co beamsPhysics in Medicine & Biology, 2011
- Comparison of 60 Co and 192 Ir sources in HDR brachytherapyJournal of Contemporary Brachytherapy, 2011
- Characterization of a fiber‐coupled luminescence dosimetry system for online in vivo dose verification during brachytherapyMedical Physics, 2009
- Optically stimulated luminescence (OSL) dosimetry in medicinePhysics in Medicine & Biology, 2008
- In-vivo dosimetry for gynaecological brachytherapy: Physical and clinical considerationsRadiotherapy and Oncology, 2005
- Monte Carlo dosimetric study of the BEBIG Co-60 HDR sourcePhysics in Medicine & Biology, 2005
- The response of thermally and optically stimulated luminescence from Al2O3:C to high-energy heavy charged particlesRadiation Measurements, 2004
- TLD, diode and Monte Carlo dosimetry of an192Ir source for high dose-rate brachytherapyPhysics in Medicine & Biology, 1995
- Prediction of severe late complications in fractionated, high dose-rate brachytherapy in gynecological applicationsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1991