Study of dielectric breakdown in liquid xenon with XeBrA: The xenon breakdown apparatus
Open Access
- 1 January 2023
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 94 (1) , 015112
Abstract
Maintaining the electric fields necessary for the current generation of noble liquid time projection chambers (TPCs), with drift lengths exceeding 1 m, requires a large negative voltage applied to their cathode. Delivering such high voltage is associated with an elevated risk of electrostatic discharge and electroluminescence, which would be detrimental to the performance of the experiment. The Xenon Breakdown Apparatus (XeBrA) is a 5-l, high voltage test chamber built to investigate the contributing factors to electrical breakdown in noble liquids. In this work, we present the main findings after conducting scans over stressed electrode areas, surface finish, pressure, and high voltage ramp speed in the medium of liquid xenon. Area scaling and surface finish were observed to be the dominant factors affecting breakdown, whereas no significant changes were observed with varying pressure or ramp speed. A general rise in both the anode current and photon rate was observed in the last 30 s, leading up to a breakdown, with a marked increase in the last couple of seconds. In addition, the position of breakdowns was reconstructed with a system of high-speed cameras and a moderate correlation with the Fowler–Nordheim field emission model was found. Tentative evidence for bubble nucleation being the originating mechanism of breakdown in the liquid was also observed. We deem the results presented in this work to be of particular interest for the design of future, large TPCs, and practical recommendations are provided.Keywords
All Related Versions
Funding Information
- U.S. Department of Energy (DE-SC0018982, DE-AC02-05CH11231)
This publication has 40 references indexed in Scilit:
- Weibull statistical analysis of area and volume effects on the breakdown strength in liquid nitrogenIEEE Transactions on Dielectrics and Electrical Insulation, 1995
- Area and volume effects on breakdown strength in liquid nitrogenIEEE Transactions on Dielectrics and Electrical Insulation, 1995
- Ionic mobilities in liquid xenonIEEE Transactions on Dielectrics and Electrical Insulation, 1994
- Gap size effect on liquid helium breakdownCryogenics, 1994
- Breakdown phenomena in liquid heliumIEEE Transactions on Electrical Insulation, 1989
- Examination of the statistics of dielectric breakdownJournal of Physics C: Solid State Physics, 1983
- Effect of an electric field on electron attachment to sulfur hexafluoride, nitrous oxide, and molecular oxygen in liquid argon and xenonThe Journal of Physical Chemistry, 1976
- Electron avalanche in liquid xenonPhysical Review A, 1974
- Electrode area effect on the electric breakdown of liquid nitrogenCryogenics, 1974
- Nonparametric Estimation from Incomplete ObservationsJournal of the American Statistical Association, 1958